LABBOX EASY 9 Operating Manual Download Page 2

Product description 

EASY 9 

may be used with glass or plastic pipettes to facilitate the pipetting of a wide range of liquids. Pipettes within 

the  volume range of 0.1 ml to 100 ml can be used. If  the instrument is correctly  used, the liquid comes into contact 
with the pipette only. 

Components 

 

Pipetting 

1.

Fit the pipette
Carefully attach the pipette by inserting it into the silicone adapter inside the nose cone. Do not use force.

2.

Fill the pipette
Press the suction bellows first whilst holding the 

EASY 9

 in a vertical position with the pipette tip down. Place the

tip of the pipette into the liquid to be aspirated. Press the aspirate lever slowly upwards allowing the liquid to fill the
pipette so that the meniscus is slightly above the required mark.
Note:
Do not allow liquid enter the 

EASY 9

. The function of the filter will be severely impaired reducing suction efficiency.

Spare filters are available see over.
The further up the lever is pressed, the stronger the suction becomes and the quicker the meniscus rises.
In the case of large pipettes the vacuum contained in the suction bellows is not sufficient to draw in all the liquid at
once. Therefore, press the suction bellows again and continue drawing up liquid.

3.

Adjust the meniscus
Wipe  the  pipette  tip  with  a  suitable  lint-free  tissue.  Press  the  dispense  lever  down  slowly  until  the  required
meniscus has been reached.

4.

Dispense the pipette
Hold the collecting vessel in an inclined position, placing the pipette tip against the collecting vessel’s wall. Press
the dispense lever down to empty the pipette.

5.

In the case of ‘blow-out’ pipettes
As  soon  as  the  meniscus  in  the  pipette  comes  to  a  standstill,  press  the  blow-out  bellows  to  expel  the  last  few
drops. To ensure a complete dispense procedure, wipe the pipette tip upwards a few millimetres along the inside
wall of the collecting vessel.

6.

‘To deliver’ pipettes
As soon as the liquid in the pipette comes to a standstill, wait for the period of time required to ensure the timed
liquid has been fully dispensed.

7.

Remove pipette
Hold the pipette at its upper end and gently twist and remove from the silicone adapter.

8.

Replace the filter and clean the adapter
If the pipette has been overfilled or the suction capacity is impaired, proceed as described below:
a) Remove the pipette
b) Unscrew the nose cone
c)

Remove the silicone adapter out of the housing by pulling it upwards. Use a wash bottle to accurately rinse the
adapter. Blow out the liquid and allow drying completely.

d) Remove the filter and replace with a new one.
e) Re-assemble

9.

Leak test
Attach a pipette and aspirate liquid to minimum of half the capacity of the pipette. Hold the pipette over a suitable
reservoir. If liquid drips from the pipette tip, dismantle the unit (see 8) and re-assemble correctly.

10. Return for repair

When returning 

EASY 9

 for repair, please  ensure the instrument has  been cleaned  and decontaminated.  Please

ask for a decontamination statement form.

Nose cone 

Silicon adapters 

aspirate and dispenser lever 

housing 

Suction bellows 

filter 

locking ring 

Blow-out bellows 