Safety
Safety Warning
If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the equipment may
be impaired.
The safety statements in this manual comply with the
requirements of the UK Health and Safety at Work Act 1974.
You must read this section to ensure the safe operation of your
instrument.
The instrument is designed to be used by properly trained
personnel and a level of knowledge is assumed.
You should adhere to the standard safety conditions in your
laboratory, in addition to the advice given in this manual.
For the correct and safe use of this instrument it is essential
that operating personnel follow generally accepted safety
procedures in addition to the safety precautions specified in
this manual.
Compressed gases present a hazard. Ensure that the
requirements for positioning and use of gas cylinders are met.
Use copper tubing for connecting gas supplies and make
sure that all connections are gas tight. Test regularly for gas
leaks. Ensure that all exits from the system are piped to a fume
cupboard.
Hydrogen Warning
Hydrogen is explosive. Do not use hydrogen unless the
application makes it essential. If you must use hydrogen, follow
the safety information supplied by your gas supplier and use
an in-line flow limiter.