6

 

 
 
 
Programming Steps and Display 

The ETC can be programmed in four simple steps using the LCD display and the three keys on the face 
of the control. 
 

Step 1 -  

   To  start  programming,  press  the  SET key once to access the Fahrenheit/Celsius 

mode.  The display will show the current status, either for degrees Fahrenheit or 
for degrees Celsius.  Then press either the up 

 or down 

 arrow key to toggle 

between the or C designation 

 

 

Step 2 - 

Press the SET key again to access the setpoint.  The LCD will display the current 
setpoint and the S1  annunciator will be blinking on and off to indicate that the 
control is in the setpoint mode.  Then press either the up 

 key to increase or the 

down 

 key to decrease the differential to the desired temperature. 

 

 

Step 3 - 

Press  the  SET key again to access the differential.  The LCD will display the 
current differential and the DIF1 annunciator will be blinking on and off to indicate 
that the control is in the differential mode.  Then press either the up 

  or  the down 

 key to decrease the differential to the desired setting. 

 

 

Step 4 -   

Press  the  SET key again to access the cooling or heating mode.  The LCD will 
display the current mode, either C1 for cooling or H1  for heating.  Then either 
press the up 

 or down 

 key to toggle between the C1 or H1 designation.  Press 

the SET key once more and programming is complete

 

Note:  The ETC will automatically end programming if no keys are depressed for a period of thirty 

seconds.  Any settings that have been input to the control will be accepted at that point

 

 

 

 

Summary of Contents for MCFB-2020

Page 1: ...s or damage sustained in transit must be made on the carrier as follows A Exterior Damage Make thorough damage notation on your delivery receipt and have driver acknowledge by signature and date Send...

Page 2: ...RUCTIONS 1 OPERATING INSTRUCTIONS 2 COMPRESSOR CONDENSER 2 SERVICE 2 Lab Research Products PRODUCT WARRANTY 3 MFB MCFB 2030 30 Models Only ETC Single Stage Electronic Temperature Control 5 Operation 5...

Page 3: ...e and is recommended on this unit POWER SOURCE The supply circuit to this cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase...

Page 4: ...with a mild soap and water solution Rinse with clear water and dry 2 Cabinet Interior Wash with 2 tbls of baking soda and 1 qt warm water solution or mild soap DO NOT USE STRONG CLEANERS SCOURING POWD...

Page 5: ...driers motor starting equipment fan assembly or any other electrical components Lab Research Products sole obligation under this warranty is limited to either repair or replacement of parts subject to...

Page 6: ...NTY CLAIMS To obtain prompt warranty service simply contact Lab Research Products at 800 648 4041 Lab Research Products shipping records showing date of shipment shall be conclusive in establishing th...

Page 7: ...ening Unless the door is left open for long periods of time this should not affect product temperature The temperature display is independent of the thermostat that is actually controlling the off on...

Page 8: ...crease or the down key to decrease the differential to the desired temperature Step 3 Press the SET key again to access the differential The LCD will display the current differential and the DIF1 annu...

Reviews: