ULTRASONIC CLEANER
5.
Trouble Shooting
Troubles
Causes
Solutions
Cleaner is not
powered.
1.
Bad connection Power Plug
2.
Power inlet Fuse is bad
3.
Fuse unit in generator is
bad.
1.
Check the power inlet
2.
Replace Fuse
3.
Request A/S
Key doesn’t
operate
1.
Key pad defective
1.
Check the key
2.
Request A/S
The temp. of
solution doesn’t
rise
1.
Heater is not set properly
2.
Heater disconnected
1.
Check the Heater Set
2.
Request A/S
Time function
doesn’t operate
1.
PROGRAM abnormal
1.
Check Time Set
2.
Request A/S
Cleaning power is
weakened
1.
Solution is old, pollution
level is high.
2.
Solution is insufficient
3.
Residue on bath bottom is
solidified
4.
The works are too much put
in the bath
1.
Replace Solution
2.
Add Solution
3.
Clean the Bath
4.
Adjust the amount of
the works until
cleaning power is
normal
Temperature
display as the ER.
1.
Temp. sensor is damaged
1.
Replace temp. sensor
2.
Request A/S
6.
Warranty
A.
The warranty period is 1 year from the date the product was purchased.
B.
During this period, damage caused by defects is repaired freely.
C.
Damage resulted from mis-operation is responsible for the end-user.