Installation
Ensure that all collars, cushions and shields are installed properly and in the correct orientation, as
shown in figure 1-2.
Figure 1-2
A/B/C/D Cushion Orientation
CUSHION
NOT USED
“A” MODELS
“B” MODELS
“C” MODELS
“D” MODELS
For best results, the centrifuge should be housed in a controlled environment. It is recommended
that the centrifuge be secured to the benchtop, table, stand etc. using the hardware supplied. Allow
for optimum working clearance around the unit. Mount centrifuge loosely with larger holes in the
tabletop, to allow room for slight movement of centrifuge. Shown in Figures 1-3. Ensure the unit is
level by adjusting the shock mounts.
Figure 1: Schematic of Mounting Bolts
Table Top View (with Zoom)
How to use Mounting Equipment