LMA-A Small Compression Load Cell INSTRUCTION MANUAL 

Thank you for purchasing LMA-A Small Compression Load Cell (hereinafter referred to 

as the LMA-A). Before using it, please read this instruction manual carefully. Also, keep 

the manual within easy reach so that you can refer to it whenever necessary. 

1. Calling the operator's attention 

The  following  cautionary  symbols  and  headlines  are  used  to  invite  the  operator's 

attention. Be sure to observe the accompanying precautions in order to safeguard the 

operator and preserve the performance of the instrument. 

Warning

Improper  handling  may  cause  serious  injury  to  the 

operator.  To  avoid  harm,  be  sure  to  observe  the 

accompanying instructions.

Caution

Cautions are given to invite the operator’s attention, in 

order  to  avoid  instrument  failure  or  mal-function.  Be 

sure to observe the accompanying instructions.

2. Safety precautions 

Caution

 Avoid water and rain. 

The  LMA-A  is  designed  to  detect  only  the  load  applied  to  the  center  axis.  Its 

installation may directly affect measurement accuracy. Therefore, carefully install 

the LMA-A, thereby avoiding a slant load, lateral load, bending moment, etc. 

The  compensated  temperature  range  of  the  LMA-A  sufficiently  covers  daily 

atmospheric temperature changes. Partial exposure to direct heat, however, must 
be avoided. If not, transient output change may occur resulting in deterioration of 
measuring accuracy. 

If an ambient temperature unavoidably exceeds the allowable temperature range, 

protect  the  whole  LMA-A  with  heat  insulation  material  in  order  to  maintain  a 

temperature within the compensated temperature range. 

When impacts or vibrations are applied to the direction of a load, the dynamic load 

shall be static load × acceleration. If the magnitude of acceleration is not known, 

rated capacity should be sufficiently large considering the dynamic load applied on 

the LMA-A. 

Do not disassemble the LMA-A. 

Do not apply shocks to or drop matters on the LMA-A. 

Calibrate the LMA-A once a year or so. (Contact KYOWA or our representative.) 

Calibration  is  necessary  whenever  the  LMA-A  is  subjected  to  an  excessive 

horizontal force component or an excessive load.

In case abnormal output value is displayed, immediately stop the measurement. If 

the LMA-A is used in the system, immediately stop the system operation.

Wire the cable with a little slack not to be pulled during use. Drawing strength of 

cable in 20N (approx.2kgf). 

The  LMA-A  is  in  a  structure  with  a  strain 

gage  bonded  on  the  upper  diaphragm  as 

shown  in  the  figure  below.  Because  the 

diaphragm is very thin, carefully handle it 

to avoid damage. 

In vibration environment, fix the  cable at its outlet and  required portions. Or, the 

cable may vibrate due to continuous load and cable may break and cause damage 

to the LMA-A. 

Make sure that the bending radius of cable is longer than 6 times of a diameter of 

the cable.

3. Mounting 

3.1 Dimensional drawing 

LMA-A-5N to 50N

                  LMA-A-100N to 1KN 

3.2 Use CC-33A adhesive to bond the load cell 

(a)  Remove  grease  and  dirt  from  the  mounting  portion  with  acetone,  methylene 

chloride, etc. If any rust, polish with sandpaper grit 120 and the like. 

(b) Apply a little drop of CC-33A as shown in the following figure and quickly spread a 

thin film on the bonding portion with a match head, etc.   

(c) Gently press and hold the LMA-A with fingers, 

etc. Since diaphragm is very thin, hold only its 

circumference with fingertips and do not press 

the center and its vicinity. 

(d) When  using  other  adhesive,  also  observe 

precautions. 

(e) When removing, insert a keen-edge cutter knife, etc. between the bottom face of 

the LMA-A and the cement position. 

Warning

Take care not to be hurt with the knives.

Caution

Do not forcedly insert a screwdriver to the circumference of the LMA-A bottom face 

or the mount may come off.

3.3 Providing a mount hole 

  5N to 50N 

:

  12.2 or higher 

100N to 1KN 

:

  20.2 or higher 

3.4 How to use the mounting base 

Attach as shown using the optional mounting base CFM. 

Mounting method using the CFM-5A          Mounting method using the CFM-100A 

NOTE:

  Cross  recessed  pan  head  screw  in  the  figure  are  included  with  each 

mounting base. 

3.5 Measuring object 

It  is  recommended  to  apply  load  through  metals  and  rigid  materials  to  have  an 

accurate measurement. 

Rigid substance, metal or the like with 

hardness HRC30 to 38 

Please use a 

cutter knife to 

remove.

Do not 

forcedly insert 

a screwdriver 

or the mount 

may come off. 

×

Strain gages 

LMA-A

Diaphragm 

Bonding Portion

Hold with 

Fingertips

CC-33A