KYORITSU KEW SNAP 2434 Instruction Manual Download Page 2

92―1509A

04―12

DISTRIBUTOR

Kyoritsu  reserves  the  rights  to  change  specifications  or  designs

described in this manual without notice and without obligations.

3-phase 3-wire system
(In  4-wire  system  with  neutral,
clamp onto all 4 wires)

Single-phase 2-wire system 
(in  3-wire  system  with  neutral,
clamp onto all 3 wires)

Load

Load

6. OPERATING INSTRUCTIONS

DANGER

●In order to avoid possible shock hazard, never make measurement

on circuits having a potential of 300VAC or greater. 

●The  transformer  jaws  are  made  of  metal  and  their  tips  are  not

completely  insulated.  Be  especially  careful  about  the  possible
shorting where the equipment under test has exposed metal parts. 

●Never  make  measurement  with  the  battery  compartment  cover

removed. 

●Keep  your  fingers  and  hands  behind  the  barrier  during

measurement.

CAUTION

●Take  sufficient  care  to  not  to  apply  shock,  vibration  or

excessive  force  to  the  jaw  tips.    Otherwise,  precisely  adjusted
Transformer Jaw tips will be damaged.

●When  a  foreign  substance  is  stuck  in  the  jaw  tips  or  they

cannot  properly  engage,  the  transformer  jaws  do  not  fully
close.  In such a case, do not release the jaw trigger abruptly or
attempt  to  close  the  transformer  jaws  by  applying  external
force.    Make  sure  that  the  jaws  close  by  themselves  after
removing the foreign substance or making them free to move.

●The  maximum  size  of  a  conductor  to  be  tested  is  28mm  in

diameter.    Accurate  measurement  cannot  be  made  on  a
conductor  larger  than  this,  because  the  transformer  jaws
cannot fully close.

●When measuring large current, the transformer jaws may buzz.

This has no effect on the instrument's performance or safety.   

6-1 Leakage Current Measurement

(1)Set  the  Range  Selector  Switch  to  the  desired  position.    Current

to measure should be within the selected measuring range. 

(2)Measuring out of balance leakage current (See Fig.1): 

Clamp  onto  all  conductors  except  a  grounded  wire.    Measured
current value is shown on the display.  

(3)Measuring earth leakage current (See Fig.2):

Clamp onto a grounded wire.Measured current value is shown on
the display.

Fig. 1 Measuring out of balance leakage current 

6-2 How to Use Frequency Selector Button

When  high  frequencies  from  such  equipment  as  inverters  are
present  in  the  circuit  under  test,  the  instrument  measures  AC
current of not only 50Hz or 60Hz of fundamental frequency but also
of these high frequencies and harmonics. 

To  eliminate  the  effect  of  such  high  frequency  noise  and  measure
AC  current  of  50Hz  or  60Hz  fundamental  frequency,  a  "high-cut"
filter  circuit  in  incorporated  into  the  instrument  which  works  when
"50/60Hz"  frequency  response  is  selected  with  the  Frequency
Selector  Button.  Cut-off  frequency  of  the  "high-cut"  filter  is  about
160Hz with attenuation characteristic of approx. -24dB/octave. 

When the Frequency Selector Button is pressed, "50/60Hz" mark is
shown on the left side of the display.  When the Frequency Selector
Button  is  pressed  again,  frequency  response  is  switched  to  WIDE
with "WIDE" mark shown on the display.  

Note:

Characteristic  of  -24dB/octiave  means  that  signal  magnitude
declines to about one sixteenth of that at the initial frequency when
frequency  doubles.  KEW  SNAP  2434  have  the  following  two
settings for the Frequency Selector Button. 

WIDE (40Hz-): Permits  measurement  of  currents  of  fundamental

frequencies  as  well  as  currents  of  high  frequencies
generated by such equipment as inverters

50/60Hz (40-approx.160Hz): Filters out high frequency currents and

measures  current  of  fundamental
frequency only

Fig.3 KEW SNAP 2434 Frequency Characteristic

Output characteristic are shown in Fig.3.

;;;;

;;;;

;;;;

;;;;

;;;;

;;;;

;;;;

100% 

50% 

20% 

10% 

5% 

5k 

1k  2k 

500

100 200

50

20

10

ー30

ー20

ー10

 0

Attenuation〔dB〕 

 10

Frequency〔Hz〕 

Percentage

Typical characteristic: 
400mA/4A Range

Typical characteristic: 

WIDE frequency response

Typical characteristic: 

50/60Hz frequency response 

high-cut filter -24dB/oct

Typical characteristic: 
100A Range

Fig.2 Measuring earth leakage current

Load

Grounded wire

Recently there has been increased use of power through inverters,
switching regulators, etc. When the high frequency noise from such
appliances  leaks  or  flows  into  the  ground  through  capacitors  not
filtering completely, the earth leakage breaker may trip even though
there is no "actual" leakage.  In such a case, the instrument do not
give  leakage  current  reading  if  "50/60Hz"  frequency  response  is
selected. 

Take  current  readings  with  the  50/60Hz  and  WIDE  frequency
responses  respectively  to  make  effective  use  of  the  Frequency
Selector Button.

6-3 Load Current Measurement

(1)Set the Range Selector Switch to the desired position. Current to

measure should be within the selected measuring range.

(2)Measuring load current (See Fig.4):

Press  the  jaw  trigger  to  open  the  transformer  jaws  and  close
them  over  one  conductor  only.Measured  current  value  is  shown
on the display.

7. OTHER FUNCTIONS

7-1 Sleep Function

This is a function to prevent the instrument from being left powered
on in order to conserve battery life.
The  instrument  automatically  enters  the  Sleep  (powered-down)
Mode about 10 minutes after the last switch operation.
To exit the Sleep Mode, press the Data Hold or Frequency Selector
Button or turn the Range Selector Switch to OFF, then to any other
position.
How to Exit the Sleep Mode:
Turn  the  Range  Selector  Switch  from 

OFF

to  another  position

with  the  Data  Hold  Button  pressed.  Then, 

P.OFF

is  shown  on

the  display.  This  disables  the  Sleep  function  and  enables
continuous use of the instrument.
To enable the Sleep function, turn the Range Selector Switch back
to 

OFF

, then to any other position.

Note:
The  instrument  consumes  small  amount  of  current  in  the  Sleep
Mode.  When  the  instrument  is  not  in  use,  make  sure  to  set  the
Range Selector Switch to 

OFF

7-2 Date Hold Function

This  is  a  function  to  freeze  the  readings  on  the  display.    When  the
Data Hold Button is pressed once, the current reading is held even
though  current  under  test  varies.  "H"  mark  is  shown  on  the  upper
right corner of the display.   
To exit the data hold mode, press the Data Hold Button again. 
Note: When the Sleep function works while the instrument is in the
data hold mode, data hold is cancelled.

Fig.4 Measuring load current

Load

8. BATTERY REPLACEMENT

CAUTION

●Do not mix new and old batteries. 
●Install  batteries  in  the  orientation  as  shown  inside  the  battery

compartment, observing correct polarity. 

●In order to avoid possible shock hazard, always set the Range

Selector Switch to the OFF position before trying to replace the
batteries

When  the  battery  voltage  warning  mark  "BATT"  is  shown  on  the
top  left  corner  of  the  LCD,  replace  the  batteries.    Note  that  the
display blanks and "BATT" mark is not shown if the batteries are
completely exhausted. 

(1)Set the Range Selector Switch to "OFF." 
(2)Loosen the battery-compartment-cover-fixing screw on the lower

back of the instrument. 

(3)Replace  the  batteries  with  two  new  R03  (UM-4)  1.5V

batteries.  

(4)Put  the  battery  compartment

c o v e r   b a c k   i n   p l a c e   a n d
tighten the screw. 

Note: U s e   a l k a l i n e   b a t t e r i e s   t o

perform measuring for long 

9. OPTIONAL ACCESSORIES

Model 8004 and 8008 (Multi-Tran)
These models help KEW SNAP 2434 to measure current greater than
1000A or to make measurement on a large bus-bar or conductor.

(1)Set the Range Selector Switch to "100A".
(2)As  shown,  open  the  jaws  and  close  them  over  the  pickup  coil

of Model 8004 or Model 8008. 

(3)Clamp on a conductor with Model 8004 or Model 8008. 
(4)Take the reading and multiply it by 10. 

Note:  Model  8004  and  Model  8008  cannot  be  used  for  leakage  current

m e a s u r e m e n t .   F o r   d e t a i l e d   s p e c i f i c a t i o n s ,   r e f e r   t o   t h e
instruction manual for Model 8004 or Model 8008.   

KEW2434

M-8004 
M-8008

Pickup Coil

Conductor

Transformer 
jaws

10

 

 

M-8004
M-8008

Max. Conductor Size

60mm in diameter

100mm in diameter

Measuring

Range

0〜1000A

Current Transformation

Ratio

10:1

Screw

Battery 
Compartment 
Cover

Batteries

Reviews: