background image

4

INSTRUMENT LAYOUT

5

PREPARATIONS FOR MEASUREMENT

1

Checking battery voltage

  

 Set the Function Selector Switch to any position other than 

the OFF position. If the marks on the display is clearly legible 

without  symbol  " 

  "  showing,  battery  voltage  is  OK. 

If  the  display  blanks  or  " 

  "  is  indicated,  replace  the 

batteries according to section 8: Battery Replacement.

NOTE

 When  the  instrument  is  left  powered  on,  the  auto-power-save 

function  automatically  shut  the  power  off;  The  display  blanks 

even if the Function Selector Switch is set to a position other 

than the OFF position in this state.

 To power on the instrument, turn the Function Selector Switch 

or press any Button. If the display still blanks, the batteries are 

exhausted.

   Replace the batteries.

2

 Make  sure  that  the  Function  Selector  Switch  is  set  to  the 

appropriate range.

  

 Also  make  sure  that  data  hold  function  is  not  enabled.  If 

inappropriate  range  is  selected,  desired  measurement 

cannot be made.

3

 Permits the measurement by placing the one test lead in the 

holster while confirming the measurement value.

#

 

WARNING

 Verify proper operation on a known source before use or taking 

action as a result of the indication of the instrument.

6-4 Diode Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "

" mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

     [ Forward-bias Diode Test ]

         Connect the red test lead to the anode and the black test 

lead to the cathode.

     [ Reverse-bias Diode Test ]

         Connect the red test lead to the cathode and the black test 

lead to the anode.

Note:

 Hold  the  clamp  sensor  in  the  holster,  while  making  the 

diode measurement

6-5 Continuity Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "   " mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

      When  the  measurement  value  is  lower  than  35±25

Ω

,  the 

instrument beeps.

Note:

 Hold  the  clamp  sensor  in  the  holster,  while  making  the 

continuity measurement

6-6 Capacitor Measurement

(1) Set the Function Selector Switch to " 

 ".

(2) Press the 

SEL

 Button. "F" mark is shown on the display.

(3) Connect the test leads to the circuit under test.

     The measured value appears on the display.

Note:

 Hold  the  clamp  sensor  in  the  holster,  while  making  the 

capacitor measurement.

6-7 Frequency Measurement

#

 

DANGER

 In  order  to  avoid  possible  shock  hazard,  never  make 

measurement  on  circuits  with  a  maximum  voltage  difference 

of  600V  or  greater  between  conductors  (300V  or  greater 

between a conductor and ground).

 Do  not  make  measurement  with  the  test  leads  connected 

to  the  circuit  under  test.  Never  make  measurement  with  the 

battery compartment cover removed.

 Do  not  make  current  measurement  with  the  test  leads 

connected to the circuit under test.

 Keep  your  fingers  and  hands  behind  the  barrier  during 

measurement.

(1) Set the Function Selector Switch to "Hz."

(2) Measuring frequency of current:

      Adjust  one  of  the  conductors  to  the  center  of  the  clamp 

sensor's arrow.

     Measured value is shown on the display.

     Measuring frequency of voltage:

      Connect  the  test  leads  to  the  circuit  under  test.  Measured 

frequency is shown on the display.

Note:

 Do not use both the clamp sensor and the test leads at 

the same time, while making the frequency measurement. 

 Hold the test leads in the holster, while making the frequency 

measurement by using the clamp sensor. 

 Hold  the  clamp  sensor  in  the  holster,  while  making  the 

frequency measurement by using the test leads.

7

OTHER FUNCTIONS

7-1 Auto-Power-Save Function

NOTE

 A small amount of current is consumed even in the power-save 

state.  Make  sure  to  set  the  Function  Selector  Switch  to  the 

OFF position when the instrument is not used.

This function helps to avoid unwanted exhaustion of the batteries 

because of leaving the instrument powered on and extend battery 

life.

The instrument automatically shifts to the power-save state about 

15 minutes after the last Function Selector Switch or other switch 

operation.

A minute before power-save mode the instrument beeps 5 times, 

finally beeps longer and then shifts to the power-save mode.

To return to the normal state:

Press  the 

  Button  or  the 

SEL

  Button  to  return  from  the 

power-save mode state to the normal state.

Note:

 Pressing the 

 Button or the 

SEL

 Button over 2 sec 

to return from the power-save mode state, the function of 

each button is also enabled.

            (ex.)  When  the  function  is  60A  and  in  the  power-save 

mode,  pressing 

SEL

  Button  over  2sec  releases  the 

power-save mode and changes from initial  AC  mode  to 

DC  mode. 

To cancel the Auto-Power-Save Function

      To  cancel  the  Auto-Power-Save  Function,  turn  the  function 

switch from OFF position to any other position with pressing 

SEL

 Button.

Note:

 When  the  function  switch  is  60A  or  120A,  pressing  A 

ZERO ADJ Button over 2 sec enables to cancel the Auto-

Power-Save Function. 

            In this case, pressing A ZERO ADJ Button over 2 sec once 

agin enables the Auto-Power-Save Function. 

To enable the Auto-Power-Save Function once again

     Turn the function switch to the OFF, and then to any position.

7-2 Data Hold Function

This  is  a  function  to  freeze  a  measured  value  on  the  display. 

Press  the 

  Button  once  to  hold  the  current  reading.  In  this 

data  hold  state,  the  reading  is  held  even  if  input  varies.  "    " 

mark is shown on the LCD. To exit the data hold state, press the 

 Button again.

Note:

 Data  Hold  Function  is  not  effective  while  making 

Continuity or Diode measurement.

  

 

 When the function is on the position where 

SEL

 Button or 

A ZERO ADJ Button is effective, pressing 

SEL

 Button or 

A ZERO ADJ Button cancels the held indication.

8

BATTERY REPLACEMENT

#

 

WARNING

 In  order  to  avoid  possible  shock  hazard,  always  disconnect 

the test leads from the circuit under test and set the Function 

Selector Switch to the OFF position before trying to replace the 

batteries.

#

 

CAUTION

Do not mix new and old batteries.

 Install batteries in the orientation as shown inside the battery 

compartment, observing correct polarity.

When the battery voltage warning mark " 

 " is shown on the 

top  left  corner  of  the  LCD,  replace  the  batteries.  Note  that  the 

display blanks and " 

 " mark is not shown if the batteries are 

completely exhausted.

(1) Set the Function Selector Switch to "OFF."

(2) Remove the instrument from the holster.

(3)  Loosen  the  battery-compartment-cover-fixing  screw  on  the 

lower back of the instrument.

(4) Replace the batteries with two new R03 (UM-4) 1.5V batteries.

(5)  Put the battery compartment cover back in place and tighten 

the screw.

 [ How to storage the clamp sensor and the test leads ]

6

HOW TO MAKE MEASUREMENT

6-1 Current Measurement

#

 

DANGER

 In  order  to  avoid  possible  shock  hazard,  never  make 

measurement  on  circuits  with  a  maximum  voltage  difference 

of  600V  or  greater  between  conductors  (300V  or  greater 

between a conductor and ground).

 Do  not  make  measurement  with  the  test  leads  connected  to 

the circuit under test.

 Never make measurement with the battery compartment cover  

removed.

 Keep  your  fingers  and  hands  behind  the  barrier  during 

measurement.

 To avoid electrical shock by touching the equipment under test 

or its surroundings, be sure to wear insulated protective gear.

#

 

CAUTION

 When handling the clamp sensor, exercise caution not to apply 

excessive shocks or vibration to the sensor.

Maximum measurable conductor size is 12mm in diameter.

Note: 

 Make  sure  that  the  test  leads  are  in  the  holster  while 

making current measurement.

  

  

 Instrument can show the value over 120A, but the safe and 

proper measurement range is limited lower than 120A.

6-1-1 DC Current Measurement

(1) The Function Selector Switch to the "60A" or "120A" position.

     ("AC" marks are shown on the top of the display.)

(2) Press the 

SEL

 Button. "   " mark is shown on the display.

(3)  Press  the  A  ZERO  ADJ  Button  to  set  the  reading  of  the 

instrument to zero.

      (Incorrect zero adjustment makes measurement errors.)

(4)  Adjust  one  of  the  conductors  to  the  center  of  the  clamp 

sensor's arrow.

      (  When  the  position  of  the  conductor  is  not  at  the  center  of 

the arrow, the error occurs.)

     Measured value is shown on the display.

Note:

 The  direction  of  current  is  plus(+)  when  the  current 

flows  from  the  upside  (A  ZERO  ADJ.  Button  side)  to  the 

underside The direction is minus(-) when the current flows 

from the underside to the upside.

  

 

  Minus "-" mark is shown at the left side of the value and 

the bar graph while making minus current measurement.

  

 

 By changing the function switch 60A or 120A to the other 

position,  the  AC/DC  mode  is  reset  to  the  default  (AC 

mode). To set the DC mode, press the 

SEL

 Button again.

  

 

 The  zero  adjustment  is  effective  only  for  the  current 

measurement. 

  

 

 After  the  zero  adjustment  the  instrument  operates  as 

follows.

    

(1) The bar graph disappears.

    

(2)  The  maximum  counts  is  changed  according  to  the 

adjusted value.

     

 (ex.)  The  maximum  count  is  6039-100=5939,  when 

ad100 counts to zero.

    

(3) 

mark is shown on the display.

    

(4)  Pressing A ZERO ADJ. Button again or 

SEL

 Button, or 

changing the function switch releases zero adjustment 

function when the zero adjustment is effective.

  

 

 Pressing the A ZERO ADJ button over 2 sec releases the 

zero adjustment function.

6-1-2 AC Current Measurement

(1) Set the Function Selector Switch to "60A." or "120A".

 ("   " mark is shown on the top of the LCD.)

(2)  Adjust  one  of  the  conductors  to  the  center  of  the  clamp 

sensor's arrow.

      (  When  the  position  of  the  conductor  is  not  at  the  center  of 

the arrow, the error occurs.)

     Measured value is shown on the display.

Note:

 Unlike  DC  current  measurement,  zero  adjustment  is  not 

necessary. There is not polarity indication either.

6-2 Voltage Measurement

#

 

DANGER

 In  order  to  avoid  possible  shock  hazard,  never  make 

measurement  on  circuits  with  a  maximum  voltage  difference 

of  600V  or  greater  between  conductors  (300V  or  greater 

between a conductor and ground).

 Do  not  make  measurement  with  the  battery  compartment 

cover removed.

 Keep  your  fingers  and  hands  behind  the  barrier  during 

measurement.

Note:

 Make  sure  that  the  clamp  sensor  is  in  the  holster  while 

making voltage measurement.

  

 

 Instrument  may  show  the  value  over  600V,  but  the  safe 

and proper measurement range is limited lower than 600V.

6-2-1 DC Voltage Measurement

(1) Set the Function Selector Switch to "   ".

  ("   " mark is shown on the LCD.)

(2) Short-circuit the tips of test leads to make the indication zero.

(3)  Connect the red test lead to the positive (+) side of the circuit 

under  test  and  the  black  test  lead  to  the  negative  (-)  side. 

Measured voltage value is shown on the display.

  When the connection is reversed, "-" is shown on the display.

6-2-2 AC Voltage Measurement

(1) Set the Function Selector Switch to "   ".

     ("   " mark is shown on the LCD.

(2) Connect the test leads to the circuit under test.

     Measured voltage value is shown on the display.
Note:

 The  displayed  value  might  be  few  digits  instead  of  zero 

even if shorting the test leads.

6-3 Resistance Measurement

#

 

DANGER

Never make measurement on circuits that are live.

 Never make measurement with the battery compartment cover 

removed.

 Keep  your  fingers  and  hands  behind  the  barrier  during 

measurement.

(1) Set the Function Selector Switch to " 

 ".

(2)  Press the 

SEL

 Button and "

Ω

" mark is shown and "   " mark 

is not shown on the display (Resistance Measurement).

     Immediately  after  setting  the  Function  Selector  Switch  to 

 ", is not necessary to operate the 

SEL

 Button.

(3)  Check that the display shows "   ". Short the test leads and 

check that the display reads about zero.

(4)  Connect  the  test  leads  to  the  circuit  under  test.  Measured 

resistance value is shown on the display. 

Note:

 When the test leads are shorted, the display may read a small 

resistance value. This is the resistance of the test leads.

  

 

 If  there  is  an  open  in  either  of  the  test  leads,  "    "  is 

shown on the display.

  

 

 The capacitive factor of the tested circuit may cause the 

fluctuation  of  the  measurement  value,  while  making  the 

resistance measurement of high value .

  

 

 Make  sure  that  the  clamp  sensor  is  in  the  holster  while 

making resistance measurement.

Adjust Button

Holster

Barrier

Clamp Sensor

LCD Display

Test Lead

Barrier

Function Selector Switch

Data Hold
Button

Range Selector
Button

A Zero

Continuity

Capacitor

Diode

Resistance

DC

AC

Clamp Sensor
Cable

Test Lead Cap

・・・・

Resistance

60A, 120A

・・・・

AC

default setting

    Button

    Button

Battery Voltage

AC / DC

Continuity

Data Hold

Bar Graph

Unit

Zero ADJ

Diode Test

Buzzer

Warning

Adjust a
conductor to
the center of
the arrow.

Correct

Incorrect

Conductor

Black

Red

Forward-bias Diode Test

Test Lead

Anode

Cathode

Red

Black

Reverse-bias Diode Test

Test Lead

Anode

Cathode

Prohibition

Battery
Compartment
Cover

Batteries

Screw

Wind the lead around the folder.

Test Leads

Clamp Sensor