INSTRUCTION MANUAL

KEW 8146/8147/8148

LEAKAGE CLAMP SENSOR 

Series

KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,

TOKYO, JAPAN

KEW8146

KEW8147

KEW8148

LEAKAGE CLAMP SENSOR

#

 DANGER

●  Never  make  measurement  on  a  circuit  in  which  the 

electrical potential exceeds AC300V.

●  Do not make measurement when thunder rumbling. If the 

instrument is in use, stop the measurement immediately 

and remove the instrument from the measured object. 

●  Do not attempt to make measurement in the presence of 

flammable gasses. Otherwise, the use of the instrument 

may cause sparking, which can lead to an explosion.

●  The transformer jaws are made of metal and their tips 

are  not  completely  insulated.  Be  especially  careful 

about  the  possible  shorting  where  the  measured 

object has exposed metal parts.

●  Never attempt to use the instrument if it's surface or 

your hand are wet.

●  Do  not  exceed  the  maximum  allowable  input  of  any 

measuring range.

#

 WARNING

●  Never  attempt  to  make  any  measurement,  if  any 

abnormal conditions are noted, such as broken case, 

and exposed metal parts.

●  Do  not  install  substitute  parts  or  make  any 

modification to the instrument. 

Return the instrument to the distributor from who you 

purchased this instrument for repair or re-calibration in 

case of suspected faulty operation.

●  Always keep your fingers and hands behind the barrier 

on the instrument to avoid the possible shock hazard.

#

 CAUTION

●  Do not step on or pinch the cord to prevent the jacket 

of cord from being damaged.

●  The  output  connector  shall  be  removed  or  connected 

without clamping a conductor. Otherwise, it may cause 

a failure.

●  Do not expose the instrument to direct sunlight, high 

temperatures, humidity or dew.

●  Never  give  shocks,  such  as  vibration  or  drop,  which 

may damage the instrument.

●  Use  a  damp  cloth  and  detergent  for  cleaning  the 

instrument. Do not use abrasives or solvents.

5

3

1

2

4

6

Transform Jaws

Trigger

Cable

Output connector

Barrier

5-1 Measuring method

(1)  Connect the Output connector to the Input terminal of 

the measuring instrument.

(2)  Press  the  Trigger  to  open  the  transformer  jaws  and 

clamp onto one conductor.

In this case, the measured conductor shall be at the 

center of the jaws.

(3) Ensure that the tips of transformer jaws are firmly closed.

5-2 Measuring method

(1)  Measuring  out  of  balance  leakage  current  (See 

Fig.1):Clamp  onto all conductors except a  grounded 

wire.

(2)  Measuring  earth  leakage  current  (See  Fig.2):Clamp 

onto a grounded wire.

3- phase 3- wire system

(In 4- wire system with neutral, 

clamp onto all 4 wires)

Fig.1 Measuring out of balance leakage current

Grounded wire

Load

Load

Load

Single- phase 2- wire system

(In 3- wire system with neutral,

clamp onto all 3 wires)

Fig.2 Measuring earth leakage current

92-1746D

3-16