19
―
7.5 Measuring current consumption (ILINE) (Only for M-6202)
This instrument can measure current consumption of the DUT. The safety
tests described in clause 7.4 and earlier must not be performed and passed.
(1) Connect the DUT to the test socket on the instrument.
(2) Set the range switch to ILINE16A range with pressing the safety lock
button on the instrument. If the safety lock button is not pressed, "no" is
displayed on the LCD and current measurement cannot be performed.
(3) Power the DUT on.
(4) Read the measured value in "A" from the LCD.
Connection for current consumption measurement
Set the range switch to ILINE16A range and power the DUT on.
WARNING
●
The maximum allowable load capacity is 16A. The main power supply to
which the test instrument is connected must be protected against
overload with a fuse or breaker. The ratings of fuse or circuit breaker
shall not exceed 16A.
●
Great attention must be paid when powering on the equipment such as
power motor.
●
The maximum throughput on current consumption measuring range
(ILINE16A) of M-6202 is 4000VA.
It shall not be exceeded.
!
Fig.18