Kusam-meco KM-CAL-805 Operation Manuals Download Page 11

16

®

®

17

  Press the key (100% / START) to store the selection; cut off
  the power again to exit from the maintenance function. 

Section Eight Performance Index

Output

Range

Output
Range

Resolution

Accuracy

DCV

0.000~22.000mA

0.001mA

Remark

20mA

-20mA

24V

0.000~-22.000mA 0.001mA

+0.05%set
 value +4uV

-

-

+0.05%set
 value +4uV

-

-

Max. load 
1 K

W

 

at 

20mA

Max. load 
1 K

W

 

at 

20mA

+10%

-

Loop
Power 
Supply

Simu-
transmitter
(absorption
 current )

Max output 
current up
to 25mA

Input

Range

Output
Range

Resolution Accuracy

Remark

+0.025%
reading
+2mV

-

-

+10%

-

Voltage

Current

Loop
Current

28V

20mA

20mA

-0.200~28.000V

-1.000~22.000mA

0.000~22.000mA

1mV

0.001mA

0.001mA

+0.02%
reading
+4uA

-

-

Input resistance

about 1M

W

resistance
about 20

W

providing 24V 
loop power

Power Supply  :  2 1.5V alkaline batteries (Lr6)

Battery life  :  about 400mA/3V under the condition of 10mA with
                      1K

W

 load

 Max permitted voltage  : 30V (between any two terminal or 
                                        between any terminal and earth ground)

Operating temperature : 0°C to 50°C

Operating relative humidity :    80% RH

Storage temperature :    -10°C to 50°C

Relative humidity for storage :     90%RH

Temperature co-efficiency : 0.1 x (designated accuracy) %/°C (5°
                                              C~18°C  28°C~40°C)

Size : 180(L) X 90 (W) X 47 (D) mm (with holster)

Weight : 500g (with holster)

Accessories : Operation instruction, a set of CF-36 industrial test
                        lead (with alligator clips attached to probes)

Safety : Certified as compliant to IES 1010 Provision (Safety
              standard issued by International  Electrotechnical
               Commission)

General Specification

v

v

v

Summary of Contents for KM-CAL-805

Page 1: ... 40 C í Power Consumption about 400mA 3V in condition of 10mA with í Dimension 180 L x 90 W x 47 D mm with holster í Weight About 500g í Power 1 5V x 2 alkaline batteries 1kW load ACCESSORIES User Manual Test lead CF 36 Clips for probe Holster Carrying case SAFETY Complies with IEC1010 safety standard issued by International Electrician Committee NEW OUTPUT FUNCTION Output Output Range Range 20 mA...

Page 2: ...l Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 K...

Page 3: ...use damage to the instrument in use The warming also point out how to avoid the accidents Caution A caution shows that if the operation does not comply with the following correct instruction it is possible to cause damage to the instrument it is possible to cause damage to the instrument in use The caution also points out how to avoid mal operation Note A note serves as a sign to remind the user t...

Page 4: ...strument Panel Layout and Functions 03 02 Iustration of LCD Display Area OUTPUT Press the key OUTPUT IN SW when the symbol OUTPUT appears it denotes that the Instrument is in a state of output INPUT Press the key OUTPUT IN SW when the symbol INPUTappears it denotes that the Instrument is in a state of input CAL When the symbol CAL appears denotes that the Instrument is in a state of calibration 0 ...

Page 5: ... and remove all test lead 2 Shake out any dirt that may be in terminals 3 Soak a new swab with alcohol Clean each terminal with the swab Replacing the Batteries This Instrument is powered by two AA batteries IECLR6 Warning To avoid electrical shock or personal injury Remove Test Leads from the Instrument before opening the battery door Close and latch the Battery Cover before using the meter Note ...

Page 6: ...Four Power on off of Instrument power on off Press the power key to turn on the power supply of the Instrument When the power is turned on the Instrument start to make self dignosis internally and the full screen is in display After this appropriate operation should be carried out Then repress it to hold it in one second and the power will be off Note To ensure correct operation of the Instrument ...

Page 7: ...pears in which 0 denotes 4mA 100 denotes 20mA Figure 5 1 4 Press the key to select the set digits of the output 5 Press the key to change the value of the set digits the value can do carry or number decrement automatically Hold the pressed key in one second and the numerical value will keep varying 25 Step Current Output 1 Connect to the Instrument as shown in Figure 5 1 2 When the key OUTPUT is p...

Page 8: ...ter is up to 30S while the auto stair step ramp pauses 5 seconds at each step 5 Press the key 100 START to start the output of the set waveform when the symbol ON appears Now repress the key 100 START again and the output will pause on a current value and the symbol OFF will appear Then press the same key again and output will continue to do the set steps from the pause value When the symbol OFF a...

Page 9: ...the Instrument Input Operation Procedure Function Operation n Operation Display Measurement Range DCI 20mA 00 000mA 025 00 mA 0 000 V 1 000 22 000 mA 31 25 112 50 mA 0 2000 28 000 v DCV 28 V 20mA Measuring DC Current 1 Insert one end of the test lead into the mA jack of the Instrument INPUT terminal and connect the other end to the output of the user s device as shown in Figure 5 3 2 Press the key...

Page 10: ...ar in the LCD Press the key and when the LCD displays AP OF the Instrument removes auto power off function And when the LCD displays AP ON the Instrument removes auto power off function Measuring DC Voltage 1 Insert one end of the test lead into the V jack of the Instrument INPUT terminal and connect the other end to the output of the user s device as shown in Figure5 4 2 Press the key OUTPUT IN t...

Page 11: ...ading 4uA Input resistance about 1MW resistance about 20W providing 24V loop power Power Supply 2 1 5V alkaline batteries Lr6 Battery life about 400mA 3V under the condition of 10mA with 1KW load Max permitted voltage 30V between any two terminal or between any terminal and earth ground Operating temperature 0 C to 50 C Operating relative humidity 80 RH Storage temperature 10 C to 50 C Relative hu...

Page 12: ...mA DCV 26V Digit meter Standard source Standard source MAX 22mA MAX 33mA MAX 33V 50ppm 0 4uA 100ppm 0 2uA 12ppm 15uV Digit meter KEITHLEY 2000 Standard resistance BZ10 100W 5520 FLUKE or equivalent Ambient Condition for Calibration Ambient temperature 23 C 1 C Relative humidity 45 to 75 RH Preheating standard instrument should be preheated to specified time put this Instrument in calibration envir...

Page 13: ...ss the key STEP AUTO and the LCD displays CAL 0 FS With the output stabilized repeat the operation of steps 5 and 6 Figure 10 1 Operating Input Calibration Operating calibrations in order of items and calibration points in Table 10 3 Item No Input Range Calibration Point 1 2 DCA 20mA DCV 28V 0 00 000mA FS 19 000mA 0 00 000V FS 19 000V Current Measurement Calibration 1 In power off state press the ...

Page 14: ... source as the value given in Table 10 3 and when the output stabilized press the key 100 START and the LCD displays SAVE denoting that the calibrated point has been stored Section Ten Notice of the Manual The present operation instruction is subject to change without notice The content of the operation instruction is regarded as correct Whenever any user find its mistakes omission etc he or she i...

Page 15: ...nd user customer of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unuse...

Reviews: