LOOP CALIBRATOR
Preliminary Data
An ISO 9001:2008 Company
G-17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA.
Sales Direct.:
022-24156638,
Tel. :
022-24124540, 24181649,
Fax :
022-24149659
Email :
Website :
www.kusamelectrical.com
An ISO 9001:2008 Company
®
Chhaya com/D/chhaya/my documents/chhaya/backup/catlog/New catlog/2015 new arrival catalogs/KM-CAL-805.cdr
Model KM-CAL-805
ELECTRICAL SPECIFICATIONS - KM-CAL-805
í
Basic Accuracy :
± 0.05%
í
Display :
4 Digit LCD display.
í
Max. Allowable Voltage :
30V..
í
Operating Temperature Range :
0 ~ 50°C.
Humidity range :
£
80% RH
í
Storage Temperature Range :
£
-10°C ~ 50°C
Humidity range :
£
90% RH
GENERAL SPECIFICATIONS
í
Temperature Coefficient :
0.1 x (designated Accuracy) % /
°C (5°C ~ 18°C, 28°C ~ 40°C)
í
Power Consumption :
about 400mA / 3V in condition of
10mA with
í
Dimension :
180 (L) x 90 (W) x 47 (D) mm (with holster)
í
Weight :
About 500g
í
Power :
1.5V x 2 alkaline batteries.
1k
W
load.
ACCESSORIES :
User Manual, Test lead CF-36, (Clips for probe),
Holster & Carrying case.
SAFETY :
Complies with IEC1010 (safety standard issued
by International Electrician Committee)
NEW
OUTPUT FUNCTION
Output
Output Range
Range
20 mA
0.000 ~ 22.000mA
DC Current
-20 mA
0.000 ~ -22.000mA
Simu-transmitter
(absorption
Current)
24 V
Loop Power
Accuracy
Resolution
0.001mA
± 0.05% set value + 4
m
A
0.001mA
± 10%
Remarks
Max. load 1K
W
at 20mA
Max. load 1K
W
at 20mA
Note:
power supply range :
5 ~ 25VDC
Max. Output current
upto 25mA
± 0.05% set value + 4
m
A
INPUT FUNCTION
Input
Output Range
Range
28 V
-0.200 ~ 28.000V
Voltage
20 mA
-1.000 ~ 22.000mA
Loop Current
Accuracy
Resolution
1mV
± 0.02% reading ± 2mV
0.001mA
Remarks
Input Resistance about
1M
W
Resistance about 20
W
Providing 24V
Loop Power
± 0.02% reading ± 4
m
A
Current
20 mA
0.001mA
0.000 ~ 22.000mA
± 0.02% reading ± 4
m
A