Kusam-meco KM -CAL 803 Operation Manual Download Page 6

07

 

 

M A X

3 0 V

3 0 V

 

M A X

 

M A X

3 0 V

IN P U T

O U T P U T

C O M

4 W

3 W

R T D

R T D

 

 

M A X

3 0 V

3 0 V

 

M A X

 

M A X

3 0 V

IN P U T

O U T P U T

C O M

4 W

3 W

R T D

R T D

Figure 6-2 3-wire connection method

Figure 6-3 4-wire connection method

 5. Press

RANG key to select resistance range or corresponding Graduation 

 of thermal resistance.

6. The  Instrument  will read the  stable measured value. The refreshing  rate  for 
measurement is nearly 1/second, and symbol'OL' may display on the screen if the 
measured value is out of range.

Section Seven Setting Functions

       Undertaking the following operation may change the auto-power-off 
function  or  the  temperature  unit  of  the  thermal  resistance  of  the 
Instrument.

1. Settings for auto-power-off function:
When the Instrument is on power-off state, press 

POWER

key for full-screen 

display,  release

POWER

key  and press

RANG

key  immediately, the 

Instrument will go into maintenance state, and symbol 'AP-XX'displaysonthescreen.
When  pressing

   

key, the Instrument may remove auto-power-off function if

 symbol 'AP- OF 'displays on the screen; The Instrument may 
restore auto-power-off function if symbol 'AP- ON 'displays on the screen. Press

4/3/2W

key to store the setting.

1. When pressing 

 

key, the Instrument will shift to the setting interface for 

temperature unit of thermal resistance, and the symbol 'TC-OC'displays on the
screen, which means Celsius degree is the temperature unit;

2.When pressing 

  

 key, the Instrument will shift to the setting interface for 

temperature unit of thermal resistance, and the symbol 'TC- OF 'displays on the 

screen, which means 

Section Eight Performance Index

Accuracy is specified for a period of one year after calibration, at 23±5ºC, with 
r e l a t i v e   h u m i d i t y   t o   7 5 % . A c c u r a c y   s p e c i f i c a t i o n s   a r e   g i v e n 
as:  ±  ([%  of  reading]  +  [number  of  least  significant  digits]) 
(“Counts”  refers  to  the  number  of  increments  or  decrements  of  the  least
significant digit).

Output function and technical index

Fahrenheit degree is the temperature unit;
Press 

4/3/2W

key to store the setting.

08

Output

function

Range Output range

Simulate 

resistance

OHM

400Ω

0.0

400.0Ω

4000Ω

0

4000Ω

Thermal 

resistance

RTD

Cu10

-10°C

250°C

Cu50

-50.0°C

150.0°C

Pt10

385

Pt100

385

Pt200

385

Pt500

385

Pt1000

385

-200.0°C

850.0°C

-200.0°C

850.0°C

-200°C

630C

-200°C

630°C

-200.0°C

630.0°C

   Resolution

Accuracy

Illustration

0.1Ω

0.05%+2

Incentive current is set as:±0.5

±3mA  When the 

incentive  current  is  set  as  ±0.1

0.5mA, 

add  an  extra  0.1Ω  to  additional  error.
The accuracy does not include lead resistance

0.05%+2

Incentive  current  is  set  as  ±0.05

±0.3mA

The accuracy does not include lead resistance

0.1°C

0.05

+6

Incentive current is set as 
±0.5

±3mAWhen the 

incentive current is set as 
±0.1

0.5mA, add an

extra 0.5°C to additional error.

Employs Pt(385) standard 
temperature The accuracy 
does not include lead
 resistance

Incentive current is set 

as± 0.05

±0.3mA

Summary of Contents for KM -CAL 803

Page 1: ...th Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital ...

Page 2: ...TENTS TITLE PAGE NO I Safety II Components and functions of meter s III Maintenance IV Power On Power Off The Meter V Output Of The Meter VI Measurement VII Setting Functions VIII Performance Index IX Test Certificate X Warranty 01 02 03 04 04 06 07 08 11 12 ...

Page 3: ...Meter periodically Do not use abrasives or solvents Note To ensure accuracy preheat for 5 minutes after power on Please contact the manufacture or dealers if the Users have higher accuracy requirement Section Two Components and functions of Meter s Panel LCD display Range selection key Function key Selection key for 2 wired 3 wired and 4 wired output selection key Input o u t p u t terminal Power ...

Page 4: ...he batteries if the meter won t be used for a long time Dirt or moisture in the terminals can affect readings Clean the terminals as follows 1 Turn the meter off and remove all test leads 2 Shake out any dirt that may be in terminals 3 Soak a new swab with alcohol Clean each terminal with the swab 2 Replacing the batteries Warning Note Replace the batteries as follows See Figure 3 1 04 Figure 3 1 ...

Page 5: ...omatically Press the key tightly the value will undertake successive change after one second 7 Press ZERO key to set the output as 000 0 C directly 5 Press key to select output setting bit 06 Section Six Measurement Warning Usage the maximum voltage allowed between the terminals and within the terminals and the ground is 30V any exceeding over this voltage may cause damage to the meter and even ma...

Page 6: ... screen which means Celsius degree is the temperature unit 2 When pressing key the Instrument will shift to the setting interface for temperature unit of thermal resistance and the symbol TC OF displays on the screen which means Section Eight Performance Index Accuracy is specified for a period of one year after calibration at 23 5ºC with relative humidity to 75 Accuracy specifications are given a...

Page 7: ... user finds its mistakes omission etc he or she is requested to contact the manufacturer The present manufacturer is not liable for any accident and hazard arising from the customer misuse or inadvertent operation The functions described in this operation instruction should not be used as grounds to apply this product to a particular purpose Section Nine Note for the Manual The present operation i...

Page 8: ...buyer or end user customer of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new...

Reviews: