01
I. SAFETY :
Terms in this manual :
WARNING :
Identifies conditions and actions that could result in
serious injury or even death to the user.
This manual contains information and warnings that must be
followed for operating the instrument safely and maintaining the
instrument in a safe operating condition. If the instrument is used
in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired. The meter is intended
only for indoor use.
CAUTION :
Identifies conditions and actions that could cause
damage or malfunction in the instrument.
Measurement Category
Measurement Category IV
is applicable to test & measuring circuits
connected at the source of the building’s low-voltage MAINS
installation. Examples are measurements on devices installed before
the main fuse or circuit breaker in the building installation.
Measurement Category III
is applicable to test & measuring circuits
connected to the distribution part of the building’s low-voltage MAINS
installation. Examples are measurements on distribution boards
(including secondary meters), circuit-breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, & equipment for industrial use & some other equipment
such as stationary motors with permanent connection to the
fixed installation
Measurement Category II
is applicable to test & measuring circuits
connected directly to utilization points (socket outlets & similar points)
of the low-voltage MAINS installation. Examples are measurements
on MAINS CIRCUITS of household appliances, portable tools &
similar equipment.
®
02
WARNING :
To reduce the risk of fire or electric shock, do not expose this
product to rain or moisture. To avoid electrical shock hazard,
observe the proper safety precautions when working with voltages
above 60 VDC or 30 VAC rms. These voltage levels pose a
potential shock hazard to the user. Do not touch test lead tips or
the circuit being tested while power is applied to the circuit being
measured. Keep your fingers behind the finger guards of the test
leads during measurement. Inspect test leads, connectors, and
probes for damaged insulation or exposed metal before using the
instrument. If any defects are found, replace them immediately.
Never attempt a voltage measurement with the test lead inserted
into the
INS./mA
or input jack that is available. Only replace
the blown fuse with the proper rating as specified in this manual.
Only use the Test Probe Assemblies provided with the equipment
or UL Listed Probe Assemblies with same rating or better. IEC
61010-031 requires exposed conductive test probe tips to be
£
4mm for CAT III & CAT IV ratings. It is commonly achieved by
permanently over-molded plastic shrouds, or by detachable Cap
shrouds for interchangeable between CAT II ratings. Refer to the
category markings on your probe assemblies as well as on the
add-on accessories, if any, for applicable rating changes.
The meter protection rating, against the users, is double insulation
per IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0, IEC/
EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and CAN/
CSA-C22.2 No. 61010-1-12 Ed. 3.0 to measurement Category III
1KV & Category-IV 600V, AC & DC. All Terminals are also rated to
such measurement categories requirements.
The meter also meets the relevant parts of EN61557 for CE
requirements, & are not certified by UL or ETL. In particular, Part 1
Ed. 2.0 General requirements, Part 2 Ed. 2.0 Insulation Resistance
& Part 4 Ed. 2.0 Resistance of earth connection & equipotential
bonding, where applicable.