Your Variable High Voltage Digital Insulation Meter has a live circuit
warning beeper. If it is connected to a live circuit, a rapid pulsating b e e p
will be heard.
DO NOT proceed to test and immediately disconnect the instrument
from the circuit. In addition your tester will display the warning
message.
5
2. SPECIFICATIONS
Test Voltage
From 500Vdc to 10KVdc
Adjustable in 500V steps
Preset Buttons
1KV, 2.5KV, 5KV, 10KV
Measuring Range
800K
W
-600G
AUTO-RANGING
Accuracy
± 5% ± 2 Digits
Power
Rechargable Battery.
Output Power Limit
1W
Voltage Regulation
Selected V 20%-5% of nominal value
Unless current limited. Meaning that if output
Current is too high, the voltage will be lowered
Automatically.
Standard Accessory :
1) Anti-leakage, Color coded, Flexible silicone
coaxial test lead with integrated guard,
Into the test probe.
2)
Manual
3)
B atteries installed.