4. OPEN, SHORT calibration:
Press “SET”, LCD will display CAL OPEn, and press
PAL/SER(ENTER), the program enters “OPEN” calibration, after
completing
the open. The LCD displays CAL SHrt, press
PAL/SER (ENTER).
The program enters Short calibration. If you
do not want to
proceed Short calibration, and press “SET” to exit.
When LCD
shows CAL SHrt, press PAL/SER (ENTER) key.
The program
enters “SHORT” calibration. And then back to
normal situation.
5. Hi/Lo Limits setting:
Press “Hi/Lo”, LCD only shows
flashing. The previous setting
Hi
will also appear and let user to do the modifications. When
you
input Lo setting value, the annunciator
is flashing. The
previous
setting Lo will also appear and let user to do the
modifications.
When the Lo setting value is greater than the Hi
setting value, the
LCD shows Err and back to the Hi setting mode.
Please to enter the
new Hi/Lo setting value.
6. TOL Hi/Lo Limits setting:
Press “TOL”, LCD shows TOL flashing. The previous setting
standard value will also appear and let user to do the modifications.
When you input +TOL setting value, the annunciators “TOL” “ ”
are flashing. The previous setting +TOL will also appear and let
user to do the modifications. When you input -TOL setting value,
the annunciators “TOL” “ ” are flashing. The previous setting
-TOL will also appear and let user to do the modifications.
7. REL setting:
Press “REL”, LCD shows D flashing. The previous standard setting
value will also appear and let user to do the modifications.
(8) MIN/MAX BUTTON
Press “Min/Max” to enter the Min/Max/Avg mode, and stop the auto
shut down function, In Addition to power and hold key, the other keys
cannot be activated. When the meter samples about 6 times, and then
beeper will emit a sound. When a new Max/Min data being recorded
and then beeper will emit 2 sounds.
Main display value (second display value) cycles through (Min/ Max
key) present value (parameter value)
→
Max value (parameter value)
→
Min value (parameter value)
→
Max value minus Min value (times
to test)
→
The Average Value's display ( times to test). The meter will
ignore and not to record if overload (“OL”) situation happens in the
comparative processes or in the capacitance range
≤
50 counts.
The Average Value's display is true average recording. It can save up
to 3000 times, the “Avg” annunciator will flash when recording times
reaching 2991 to 3000 times. When recording over 3000 times the
Avg will stop to record and it will also display the average value on
the LCD. The meter continues to record the Max/Min value.
Under the Max/Min function, press “HOLD” key to stop recording
temporarily but store the former recording value in advance, press
“HOLD” again to go on recording. In order to prevent any mistakes or
losses taken, you have to press 2 sec to exit the Max/Min function and
cancel the original recordings.
(9) SET
1.
The “SET” can only be activated only before you have not used any
other functions.
2.
Press “SET” to enter the SET mode, and change to manual range
mode automatically.
3.
While in the SET function, the main display is cleared, the second
display shows “SET” annunciators, LCD shows
Δ
, TOL,
,
,
flashing. There are only 5 keys that you can use “Power”, “SET”,
“REL”, “Hi/Lo”, “TOL” in this moment.