Kusam-meco KM 3600 Operation Manual Download Page 6

“KUSAM-MECO”

AN ISO 9001:2008 COMPANY

“KUSAM-MECO”

AN ISO 9001:2008 COMPANY

07

06

2.  Button “°C/°F/°K”

Press the °C/°F/°K button to cycle through temperature scale, 
°C, °F and °K.

3.  Type Selection (thermocouples)

Press the “type” button and the selected symbol will blink which 
means in the setting mode.
Press “ 

}

 ” button to make right shifts to cycle through

                  K

g

J

g

T

g

E

g

R

g

S

g

N

g

L

g

U

g

B

g

C

g

mV.

Press “ 

|

 ” button to make left shifts to cycle through

                  K

g

mV

g

C

g

B

g

U

g

L

g

N

g

S

g

R

g

E

g

T

g

J

Press the “type” button again to choose the selected thermocouple.

4.  Backlight “     ” button

Pressing the button less  than  two seconds to turn on and pressing
the button  again  less than two  seconds  to turn off the backlight in 
the LCD.  It will turn off after thirty minutes without operation.

Auto Power Off

It is in the APO mode when the meter is turned on and will turn off
the meter without operation for fifteen minutes.  Pressing the “     ” 
button over two seconds to cancel the function.  And pressing the
“       ” button again to activate the APO function.

5.  “ REL” button

The  relative  value  function  can  be used for comparing the saved 
reference value with other measurements.  Press  the “ REL” button
less than to store the current  measurement as the reference value,
and press the “ REL” over two seconds to disable the function.

6.  “ HOLD” button

When HOLD mode is selected, the thermometer  holds  the present 
readings and stops all further measurements.  To activate the  data
hold mode, press the HOLD button, and “HOLD” is displayed on the
LCD.  Pressing  the HOLD button again cancels the function, &  the
instrument   will  automatically  resume  measurements.  When  the 
Hold  key  is  activated,  it  will  stop functions of the other entire key 
except Power & Backlight.

7.  “ T1/T2 SOURCE” button

Pressing T1/T2 SOURCE to cycle through T1, T2 and SOURCE.  In
the main display the blinking digit is the one to be adjusted, you can
push the “ 

|

  

}

 ” button to make right or left shifts to the desired 

position.  When incrementing to the utmost range of the selected 
thermocouples, the LIMIT will show on the display.
SOURCE is to provide the output parameter settings.  There are ten
individual temperature setting points in group 0, which can be set at
your desired output point.  Use “ 

|

  

}

 ” to shift the desired digit to be 

adjusted and rotate the knob to increase or decrease the values you
want to set.  Press the T/C OFFSET to save the settings.

8.  “ OFFSET” button (Thermocouple offset adjust)

When the main display input is T1 or T2, and socket thermocouple is
connected.    Press  T/C  OFFSET  over  two  seconds  the  SET 
annouciator  will  appear  on  the  right  side  of  display  and  enter  the 
offset  adjustment  mode.    And  the  blinking  digit  is  the  one  to  be 
adjusted.  Rotate the knob to the right increasing the values, to the 
left side decreasing the values.  The maximum range of the knob is
±5  centigrade.    When  turning  to  the  utmost  range,  it  will  appear 
LIMIT symbol on the left side of the display and means that there is
no  further  incrementing  of  the  offset.    Press  the T/C  OFFSET  over 
two seconds to save the settings.

9.  “ MIN / MAX ” button

Press  MIN / MAX  button to enter the MIN/MAX recording mode and
REC  shows  on  the  display.  The  beeper emits a tone when a new 
minimum  or  maximum  measurement  is recorded.  Press the MIN /
MAX button again to rotate through the current readings :
MAX : The highest measurement recorded.
MIN  : The lowest measurement recorded.
MAX-MIN : The difference of the highest and the lowest 
                   measurement.
AVG. : The average values of the measurements.
Press MIN/MAX button over two seconds to exit the function.

Summary of Contents for KM 3600

Page 1: ... o o o o 200 C to 1372 C or 328 F to 2501 F o o o o 210 C to 1200 C or 346 F to 192 F o o o o 250 C to 400 C or 418 F to 752 F o o o o 250 C to 1000 C or 418 F to 1832 F o o o o 0 C to 1767 C or 32 F to 3212 F o o o o 0 C to 1767 C or 32 F to 3212 F o N TYPE 0 1 o L TYPE 0 1 o U TYPE 0 1 o B TYPE 1 o C TYPE 1 o o o o 200 C to 1300 C or 328 F to 2372 F o o o o 200 C to 900 C or 328 F to 1652 F o o ...

Page 2: ...l Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Function Generator Phasing Sticks Battery Tester Waterproof Pen Testers Solar Power Meter EMF Detector Wood Paper Grain Moisture Meter Transistorised Electronic Analog Digital Insulation Resistance Testers upto 10 K...

Page 3: ... in the leads especially near the connector CAUTION INTRODUCTION This instrument is a 4 digit compact sized portable digital calibrator thermometer designed to use external K J T E R S N L U B C type thermocouples as temperature sensor The thermometer featres a dual thermocouple input an adjustable T C offset The thermocouples types comply with the N I S T Monograph 175 revised to ITS 90 standard ...

Page 4: ...e value of T1 T2 or output setting The second displays T1 or T2 readings and the third T1 T2 and groups settings There are three different display areas Main second and GENERAL SPECIFICATIONS Battery Battery life is about 17 5 hours when used with a carbon zinc battery Standard 9V battery NEDA 1604 IEC 6F22 006P Temperature Scale Celsius or Fahrenheit user selectable ELECTRICAL SPECIFICATIONS Meas...

Page 5: ...o 50 C 32 F to 64 F and 82 F to 122 F Top Side 1 Output 1 2 Output 2 3 Input T1 4 Input T2 5 RS232 optical interface RS232 OPTICAL INTERFACE OUTPUT OUTPUT T1 T2 1 2 4 3 5 C F K APO 2S TYPE REL HOLD MIN MAX T C OFFSET LCD DIGIT DISPLAY T1 T2 SECOND MAIN T1 T2 SOURCE CALIBRATOR THERMOMETER APO HOLD T1 T2 SET T1T2 T1 T2 COM REC AVG MAX MIN OUT REL LIMIT Third Second Main 1 4 7 9 8 6 5 3 2 1 Power But...

Page 6: ...other entire key except Power Backlight 7 T1 T2 SOURCE button Pressing T1 T2 SOURCE to cycle through T1 T2 and SOURCE In the main display the blinking digit is the one to be adjusted you can push the button to make right or left shifts to the desired position When incrementing to the utmost range of the selected thermocouples the LIMIT will show on the display SOURCE is to provide the output param...

Page 7: ...n set 10 step Group 1 to 9 use software setting press T C OFFSET exit group set mode Group out In the OUT mode press T C OFFSET less than two seconds Rotary knob to select step Press T C OFFSET exit group out mode WARNING To avoid possible electrical shock disconnect the thermocouple connectors from the thermometer before removing the cover Battery Replacement Periodically wipe the case with a dam...

Page 8: ...stall into a different folder click the Browse button and select another folder You can choose not to install 515R by clicking Cancel to exit the installation wizard Destination Folder C Program Files 515R Browse 515R Setup Next Cancel Back Click Next Ready to install the Application Click Next to begin installation Click the Back button to reenter the installation information or click Cancel to e...

Page 9: ... MIN AVG Hi LO 24 3 23 5 23 9 200 0 0 0 HOLD Clear T1 00024 3 T2 ALARM Disabled MAX MIN AVG Hi LO 26 1 25 6 25 8 1 0 0 0 HOLD Clear T1 00024 1 Channel Subtract ALARM Disabled MAX MIN AVG Hi LO 1 8 1 9 1 8 1 0 0 0 HOLD Clear Su 00001 8 T1 T2 STATE Com I Initial OK 11 38 40 PM 515R File Control Help C K Type C K Type C K Type 1 T1 T2 Channel Subtract NONE T1 T2 T2 T1 ALARM function Disable Hi enable...

Page 10: ...Calibrate Out v1 Deg ºC Unit K Type Type State Range 200 0 1372 0 0 1 2 3 4 5 6 7 8 9 D5 0 1 2 3 4 5 6 7 8 9 D6 K Type Value Send T1 OS 00036 0 C K Type 0 1 2 3 4 5 6 7 8 9 D1 0 1 2 3 4 5 6 7 8 9 D2 0 1 2 3 4 5 6 7 8 9 D3 0 1 2 3 4 5 6 7 8 9 D4 Calibrate Out Rec v2 Export Both Display drawing T1 and T2 T1 Display drawing T1 T2 Display drawing T2 State Value Setting Interval Start Log File type txt...

Page 11: ... Read EXIT ID Unit 6 Meter Temperature Unit ºC ºF ºK Type K J T E R S N L U B C mV Group Group Item Group 0 to Group 9 Unit ºC ºF ºK Type K J T E R S N L U B C mV Step Item Step 1 to Step 10 Edit Default 515R setting values Clear Clear the current group Clear All Clear all group s value 7 File Load From PC load the stored value Save Save the setting value into PC Setup Set State 10000 Meter Setup ...

Page 12: ...IT MUMBAI TEST CERTIFICATE TEMPERATURE CALIBRATOR This Test Certificate warrantees that the product has been inspected and tested in accordance with the published specifications The instrument has been calibrated by using equipment which has already been calibrated to standards traceable to national standards ISO 9001 REGISTERED QC PASS KUSAM MECO MODEL NO ________________ SERIAL NO ___________ DA...

Page 13: ... of charges provided the to and fro freight charges are borne by you This warranty extends only to the original buyer or end user customer of a KUSAM MECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by acciden...