I.SAFETY INFORMATION

● Read the following safety information carefully before 
   attempting to operate or service the meter.
● To avoid damages to the instrument, do not exceed the 
   maximum limits of the input values shown in the technical 
   specification tables.
● Never measure current while the test leads are inserted into 
   the input jacks.
● Do not use the meter or test leads if they lock damaged. Use
   with extreme caution when working around bare conductors 
   or bus bars.
● Accidental contact with the conductor could result in electric 
   shock.
● Use the meter only as specified in this manual; otherwise, the
   protection provided by the meter may be impaired.
● Read the operating instructions before use and follow all 
   safety information.
● Be cautious when working with voltages above 60V DC or 
   30V AC RMS. Such voltages pose a shock hazard.
● Before taking resistance measurements or testing acoustic
   continuity. disconnect circuit from main power supply and all
   loads from circuit.

II)  FEATURES:

 

Measurement CAT II 600V AC/DC

 

Pollution Degree 2 

 

Altitude upto 2000 meters


■ Indoor use only
■ Relatively humidity 80% max.

Maintenance & Clearing:

1. Repairs or servicing not covered in this manual should only be
    performed by qualified personnel.
2. Periodically wipe the case with a dry cloth. Do not use 
    abrasives or solvents on this instrument.

Maximum voltage between any terminal and earth ground 750 Vrms

02

GENERAL SPECIFICATIONS:

í

 

Display :

 3-5/6 digits liquid crystal display (LCD) 

     

Max.  Rading 5999. Automatic indication of 

     functions and symbols.

 

Cables 

Φ

36mm 

í

J

aw Size :

 

í

 

Maximum Voltage between any terminal & 

     Earth ground 750 Vrms.

í

 

Operating Temperature : 

0~40°C

 

Operating Principle :

 Dual slope integration

í

 

LCD will show a “OL” in the highest

í

 

Over Range Indication :

     position accompanied.  (except for ranges of 1000A AC, and 
     1000VDC or 750VAC)

 

 The     is displayed when the battery

í

 

Low Battery Indication :

     voltage drops below the operating voltage.

2.5times/sec (Digital Display).

í

 

Sampling rate : 

 

 60hrs approx.(Alkaline)

í

Battery Life :

 Automatic polarity “-” display for negative input

í

 

Polarity :
Auto Power Off :

  The meter is powered off 15 minutes 

í

 

     later after the last operating was made.  To bringe back 
     the display please turn rotary switch to more positions 
     or push any button

 

0°C to 40°C (R.H.<80%)

í

 

Operating Temperature & Humidity :

 -

10°C to 60°C (R.H.<80%)

í

 

Storage Temperature & Humidity :

1.5V  AA x 2  Alkaline Battery.

í

 

Power Supply : 

 

228(L) x 75(W) x 36(H)mm

í

 

Dimension :

 

Approx. 465 gms. (including battery).

í

 

Weight :

ACCESSORIES :

Test leads set, Users Manual, Battery & Carrying Case.

Summary of Contents for KM 3060-T

Page 1: ...Meter Digital Lux Meter Network Cable Tester Power Factor Regulator Earth Resistance Tester Digital Panel Meters DC Power Supplies High Voltage Detector Calibrators Gas Analysers Frequency Counter Fun...

Page 2: ...overed in this manual should only be performed by quali ed personnel 2 Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instrument Maximum voltage between any termi...

Page 3: ...current error is speci ed within the largest circle which can be drawn inside the jaw 2 2 Measurement Speci cations 04 RESISTANCE Range Overload Protection 250V RMS Accuracy Resolution 600 W 1 2 rdg 2...

Page 4: ...on In the backlight on status press the key over 2 seconds the backlight will be o 06 IV OPERATING INSTRUCTION 4 1 Measurement Procedure operating instructions 1 Make sure that the selected range is s...

Page 5: ...shock hazard or damage to the instrument 08 3 Connect the red test to the anode side and black test lead to the cathode side of the diode being tested 4 Read forward voltage Vf value on LCD If the tes...

Page 6: ...is below 50 it will indicate by a continuous beeping NOTE Continuity test is available to check open short of the circuit 10 V MAINTENANCE 5 1 Battery Replacement WARNING To prevent electrical hazard...

Page 7: ...patch of goods In case any defect occurs in functioning of the instrument under proper use within the warranty period the same will be rectified by us free of charges provided the to and fro freight c...

Reviews: