An ISO 9001:2008 Company
®
G-17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA.
Sales Direct.:
022 -24156638,
Tel. :
022-24124540, 24181649,
Fax :
022 - 24149659
Email :
Website :
www.kusamelectrical.com
An ISO 9001:2008 Company
®
All Specifications are subject to change without prior notice
D:\Chhaya\My Documents\Chhaya\backup\catlogs\NEW CAT\2013-14\KM 2775 revised 28 march 15.cdr
11 FUNCTIONS 34 RANGES
Model KM 2775
CA
T III 600V
CE
Approved
SAFETY :
?
Withstand Voltage :
6000V
?
IEC/EN61010-1 EN61010-2-32 CAT III 600V
SPECIAL FEATURES :
?
Auto Ranging
?
Full Automatic measurement
?
Range change function
?
Data Hold function freezes the reading
?
Continuity & Diode Measurements
?
Auto Power Off
?
Low Battery Indication
GENERAL SPECIFICATIONS :
í
Sensing :
Average Sensing.
í
Jaw opening size :
55mm Max.
í
Display :
3¾ digits 4000 counts LCD display
í
Sample Rate :
Approx. 2 times per second
í
Response Time :
Approx. 1 second
í
Low Battery :
“ ” sign appears on the display when the battery Voltage drops below accurate
operating level.
í
Over range indication :
“O.L” indicated
í
Operating Principle :
Dual slope integration
í
Storage Temperature :
-20°C
to 60°C, <80% R.H.
í
Temperature & Humidity for Guaranteed accuracy @ -10°C to 50°C at <80% R.H.
í
Power Supply :
Standard 9V battery
í
Dimension :
260(L) X 95(W) X 43(D) mm
í
Weight :
Approx. 515
gm (battery included)
ACCESSORIES :
Test leads, Carrying Case, Battery installed,
User's Manual.
ELECTRICAL SPECIFICATIONS - KM 2775
Accuracy is ± (% reading number of digits ) or otherwise specified, at 23°C ± 5°C & less than equal to 80% R.H.
DC CURRENT
400 A
2000 A
Range
±(1.5%rdg + 4dgts)
Resolution
Accuracy
0.1 A
1 A
±(1.5%rdg + 4dgts)
RESISTANCE
400
W
4 K
W
40 K
W
400 K
W
4 M
W
40 M
W
Range
Resolution
0.1
W
1
W
10
W
100
W
1 K
W
10 K
W
±(1.5%rdg + 3dgts)
Accuracy
±(1.5%rdg + 3dgts)
±(2.0%rdg + 4dgts)
±(1.5%rdg + 3dgts)
±(1.5%rdg + 3dgts)
±(1.5%rdg + 3dgts)
Overload Protection :
500V AC rms or 500V DC
FREQUENCY
4 KHz
20 KHz
Range
Resolution
1 Hz
10 Hz
Accuracy
±(0.3%rdg + 2dgts)
±(0.3%rdg + 2dgts)
Overload Protection :
500V AC rms or 500V DC
Trigger level :
0.2V
AC VOLTAGE
Range
Resolution
Accuracy
400 mV
4 V
40 V
400 V
600 V
0.1 mV
1 mV
10 mV
100 mV
1 V
±(2.0%rdg + 3dgts)
±(1.5%rdg + 3dgts)
±(1.5%rdg + 3dgts)
±(1.5%rdg + 3dgts)
±(1.5%rdg + 3dgts)
Overload Protection :
600V AC RMS
Frequency Response :
0 ~ 400mV at 40Hz ~ 100Hz
4V ~ 600V at 40Hz ~ 500Hz
DC VOLTAGE
400 mV
4 V
40 V
400 V
600 V
Range
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
±(1.0%rdg + 3dgts)
Accuracy
±(1.0%rdg + 3dgts)
±(1.0%rdg + 3dgts)
±(1.0%rdg + 3dgts)
±(1.0%rdg + 3dgts)
Overload Protection :
600V AC RMS
Frequency Response :
40Hz ~ 1kHz
AC CURRENT
400 A
2000 A
Range
±(2.0%rdg + 4dgts)
Resolution
Accuracy
0.1 A
1 A
±(2.0%rdg + 4dgts)
Overload Protection :
500V AC rms or 500V DC
CONTINUITY TEST
400
W
Range
Audible Threshold
Less than 25
W
DIODE TEST
2 V
Range
Resolution
±(1.5%rdg + 3dgts)
Accuracy
1 mV
CAPACITANCE
4 nF
(30p~4nF)
40 nF
400 nF
4
m
F
40 F
m
400 F
m
4 mF
Range
Resolution
1 pF
10 pF
100 pF
1 nF
10 nF
100 nF
1 F
m
Overload Protection :
500V AC rms or 500V DC
±(3.0%rdg + 5dgts)
Accuracy
±(2.0%rdg + 5dgts)
±(2.0%rdg + 5dgts)
±(2.0%rdg + 5dgts)
±(2.5%rdg + 5dgts)
±(2.5%rdg + 5dgts)
±(2.5%rdg + 5dgts)
3¾ DIGIT 2000A AC/DC
DIGITAL CLAMPMETER
WARNING
Please do not change to any other DCV or Ohm range when you are
inputing 600V AC/DC. Make sure that you have turned off the input
voltage and then you can change to another range for testing.