1. SAFETY PRECAUTIONS :

Electricity  can  cause  severe  injuries  even  with  low  voltage  or 
currents.  Therefore  it  is  extremely  important  that  you  read  the 
following information before using your digital RCCB / ELCB tester.
1.1 This   instrument   must   only   be   used  and  operated  by  a 
      competent  trained  person  and  in  strict  accordance with the 
      instructions.  We  will  not  accept  liability  for  any  damage or 
      injury  caused  by  misuse  or non compliance with instructions
      and safety procedures.
1.2 Never open your digital RCCB / ELCB tester except for battery
      replacement. (See Battery Replacement section).
1.3 Always  inspect  you digital RCCB / ELCB tester and test leads
      before  use  for  any  sign  of  abnormality  or  damage.  If  any
      abnormal  conditions  exist  (broken  test leads,  cracked case,
      display faulty etc.) do not attempt to take any measurement or
      use the tester. Return your digital RCCB / ELCB tester to your
      nearest Distributor for service.
1.4 Never  replace  the  protective  fuse  with  any  other  than  the 
      specified or approved equivalent.
1.5 Your  digital  RCCB / ELCB tester has been designed with your 
      safety   in  mind.  However, no design  can  completely  protect 
      against incorrect use. Electrical circuits can be dangerous and/
      or lethal when a lack of caution or poor safety practice is used.
      Be careful of the voltage presented above 24V as  these  pose 
      a shock Hazard.
1.6 Pay attention to cautions and warnings which will inform you of
      potentially dangerous procedures.
1.7 Rated environmental conditions :

(1) Indoor use.
(2) Installation Category III.
(3) Pollution Degree 2.
(4) Altitude up to 2000 Meters.
(5) Relative Humidity 80% Maximum.
(6) Ambient Temperature 0-40ºC.

2. SPECIFICATIONS :

Current Settings

Current Selection
Phase Start Selection
0º and 180º
Over-Temperature Protection
Wiring Correctness indication
Trip Indicator
Phase Polarity Trip Indicator
Operating Voltage (L-E) 50Hz or 60Hz
1812EL-A
1812EL-B
1812EL-C
1812EL-D
Voltmeter (L-E)
Timer Resolution
Timer Accuracy
Current Accuracy
Voltmeter Resolution
Voltmeter Accuracy
Operating Temperature
Storage Temperature
Battery

Current Specified at

3ma, 5mA, 7mA, 10mA,
15mA, 20mA, 30mA,35mA,
50mA, 100mA, 125mA, 
150mA, 175mA,
250mA, 300mA, 375mA,
500mA.
Rotary Switch 
Referenced to Earth
Yes
Yes (3 sensors)
Yes (LEDs)
Yes (LCD)
Yes (LCD)
Referenced to Earth

240Vac
230Vac
220Vac
110Vac
20Vac-280Vac
1mS (max time= 2.999S)
±2% rdg ±2mS
±5% ±1mA
1Vac
±2% ±1Vac
-5ºC to 45ºC
-10ºC to 85ºC

´

 AA batteries

Bat OK Led = Vbat >7.5v
Measure battery voltage at 
start up
Operating voltage

Summary of Contents for KM 1812EL

Page 1: ...W Mumbai 400 015 INDIA Sales Direct 022 24156638 Tel 022 24124540 24181649 Fax 022 24149659 Email kusam_meco vsnl net Website www kusamelectrical com www kusam meco co in MODEL KM 1812EL OPERATION MANUAL AN ISO 9001 2008 COMPANY DIGITAL ELCB TESTER ...

Page 2: ...nt Layout Lid Instruments RCD Test Time Delay Preparation For Measurement Battery Replacement Servicing And Calibration Test Certificate Warranty Specifications Fuse Replacement Cleaning And Storage TITLE EARTH NEUTRAL LINE MULTIFREQUENCY 50Hz 60Hz C Battery OK LIT DO NOT PROCEED LIT WIRING CORRECT N E P N P E AUTO POWER OFF 0º 180º OFF 10 15 20 30 35 50 100 125 150 175 250 275 300 375 500 3 5 7 P...

Page 3: ...otect against incorrect use Electrical circuits can be dangerous and or lethal when a lack of caution or poor safety practice is used Be careful of the voltage presented above 24V as these pose a shock Hazard 1 6 Pay attention to cautions and warnings which will inform you of potentially dangerous procedures 1 7 Rated environmental conditions 1 Indoor use 2 Installation Category III 3 Pollution De...

Page 4: ... testing 16 Characters Liquid Crystal Display 4 CONNECTIONS Line Hot Probe Connection Neutral Probe Connection Earth Ground Probe Connection LINE NEUTRAL EARTH apply only Operating Voltage 10 50Hz or 60 Hz 5 INSTRUMENT LAYOUT EARTH NEUTRAL LINE MULTIFREQUENCY 50Hz 60Hz C Battery OK LIT DO NOT PROCEED LIT WIRING CORRECT N E P N P E AUTO POWER OFF 0º 180º OFF 10 15 20 30 35 50 100 125 150 175 250 27...

Page 5: ...t the display will show T 2 999S and Hold OVER meaning the RCD did not trip below 19 999S The tripping point is out of the Range of the instrument or RCD faulty TEST PROCEDURE 1 Insert the leads into Instrument 2 Switch Instrument ON 3 Select the current using the rotary switch 4 Select positive 0º or negative 180º edge to start 5 Connect the tester to the circuit under test 6 The tester measures ...

Page 6: ...eplace the battery cover 9 FUSE REPLACEMENT 1 The fuse is located in the battery compartment To replace the fuse proceed as per Battery Replacement Section to open the battery cover then remove and replace the fuse located on the side of the batteries 2 Make sure to place the fuse protection cover small rubberized fuse cover 3 Only replace with the same specification fuse 1A fast blow 10 SERVICING...

Page 7: ...ECO authorized dealer This warranty does not apply for damaged Ic s fuses burnt PCB s disposable batteries carrying case test leads or to any product which in KUSAM MECO s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling KUSAM MECO authorized dealer shall extend this warranty on new and unused products to end user custom...

Reviews: