10
KUSAM-MECO April 2011
®
Capacitance
Range
Resolution
Accuracy
1)
60.00 nF
600.0 nF
±(2.0% rdg + 5 dgts)
6.000
m
F
60.00
m
F
Accuracies with film capacitor or better
Temperature Coefficient :
0.25 x (specified accuracy)/
°C @(0°C--18°C or 28°C – 40°C)
1)
2)
600.0
m
F
2000
m
F
±(3.5% rdg + 5 dgts)
2)
±(4.0% rdg + 5 dgts)
2)
~Hz Line Level Frequency
Accuracy :
0.03% + 4d
1)
VFD sensitivity linearly decreases from 10% F.S. @ 200Hz
to 40% F.S. @ 420Hz
Function
6 V
60 V
600 V
VFD 6 V
1)
VFD 60 V
1)
VFD 600 V
1)
Sensitivity
(Sine RMS)
1 V
6 V
60 V
1 V ~ 2 V
6 ~20 V
60 V ~ 200 V
6 A ~ 20 A
60 A ~ 200 A
VFD 60 A
1)
VFD 600 A
1)
Range
10Hz ~ 1999Hz
10Hz ~ 1999Hz
10Hz ~ 1999Hz
10Hz ~ 420Hz
10Hz ~ 420Hz
20Hz ~ 420Hz
20Hz ~ 420Hz
10Hz ~ 420Hz
Diode Tester
Range
Test Current
(Typical)
1.000V
0.56mA
<1.8V DC typical
Open Circuit
Voltage
Accuracy :
1.0% + 3d
10 pF
100 pF
1 nF
10 nF
100 nF
1
m
F
11
KUSAM-MECO April 2011
®
ACA Current(Clamp on)
1)2)3)
Accuracy
Resolution
Range
1)
Add 12d to specified accuracy while reading is at 1% to
10% of range.
2)
Induced error from adjacent current carrying conductor :
<0.5A/A
3)
Specified accuracy is from 1% to 100% of range and for
measurements made at the jaw center. When the conductor is
not positioned at the jaw center, position errors introduced are:
Add 1% to specified accuracy for measurements made WITHIN
jaw marking lines (away from jaw opening)
Add 4% to specified accuracy for measurements made beyond
jaw marking lines (toward jaws opening)
±(1.8%rdg + 3dgts)
50Hz – 400Hz
60.00 A
600.0 A
10 mA
100 mA
NON-CONTACT EF-Detection
Typical Voltage
20V (tolerance : 10V ~ 36V)
55V (tolerance : 23V ~ 83V)
110V (tolerance : 59V ~ 165V)
220V (tolerance : 124V ~ 330V)
440V (tolerance : 250V ~ 600V)
Bar-Graph
Indication
-
- -
- - -
- - - -
- - - - -
Indication :
Bar-graph segments & audible beep
tones proportional to the field strength
Detection Frequency :
50/60Hz
Detection Antenna :
Top side of the stationary jaw
Probe-Contact EF-Detection:
For more precise
indication of live wires, such as distinguishing
between live and ground connections, use the
Red (+) test probe for direct contact measurement.