![Kusam-meco KM 111M Operation Manual Download Page 5](http://html1.mh-extra.com/html/kusam-meco/km-111m/km-111m_operation-manual_2011825005.webp)
-5-
®
®
-6-
®
®
Accuracy is ± (% reading number of digits) or otherwise
specified, at 23°C ± 5°C & less than 75% R.H.
DC VOLTAGE
Range
Resolution
Accuracy
400.0 mV
4.000 V
40.00 V
400.0 V
600 V
100.0
m
V
1 mV
10 mV
100 mV
1 V
± (0.3% rdg + 4 dgts)
± (0.5% rdg + 3 dgts)
± (0.5% rdg + 3 dgts)
± (0.5% rdg + 3 dgts)
± (1.0% rdg + 4 dgts)
NMRR :
>50dB @ 50/60Hz
CMRR :
>120dB @ DC, 50/60Hz, Rs = 1k
W
Input Impedance :
10M
W
, 30pF nominal (1000M
W
for 400.0mV range)
AC VOLTAGE (50Hz ~ 500Hz)
Range
Resolution
Accuracy
4.000 V
40.00 V
400.0 V
600 V
1 mV
10 mV
100 mV
1 V
± (1.5% rdg + 5 dgts)
± (2.0% rdg + 5 dgts)
CMRR :
>60dB @ DC to 60Hz, Rs = 1k
W
Input Impedance :
10M
W
, 30pF nominal
± (1.5% rdg + 5 dgts)
± (1.5% rdg + 5 dgts)
True RMS accuracy of ACV is specified from 5% to 100% of range, &
ACA is specified from 10% to 100% of range, or otherwise specified.
Maximum Crest Factor <1.75 : 1 at full scale & <3.5 : 1 at half scale,
and with frequency spectrums, besides fundamentals, fall within the
meter specified AC bandwidth for non-sinusoidal waveforms.