Kurz Instruments Inc. 

 

December 13, 2007 

 

 
360210-AH 

Rev.A 

  504/534FTB 

Installation 

  AH 

700494-) frame or pipe stand are sufficient.  It is important to know that the sensor serial 
number must be matched with the same serial number electronics unit.  These two 
parts are not interchangeable unless the electronics configuration file is matched with 
the sensor serial number.

 

 

 

Power and Signal wires.  2x 
¾” conduit ports 

Conduit seal for 
Ex applications 

Sensor Wire 
Port: ¾” NPT 

 

Figure AH-4. Examples of Unistrut and Pipe mounting of the TS configured electronics 
using the optional mounting kit, 700494-. 

 

Field Wiring 

 
Proper wiring installation of the MFT B-Series flow transmitters may include some or all 
the flowing issues: 
 

  Safety Grounding and Explosion-Proof enclosure connections. 

  Water ingress protection 

 

Summary of Contents for 504FTB

Page 1: ...nsmitters are generally mounted with a flange pipe fitting but thread is sometimes used on smaller line sizes See the product brochures DCN 367524 for the 504FTB and DCN 360723 for the 534FTB for Kurz...

Page 2: ...onger if its electronics head is vertical up tilting down to no more than to 8 AM or 4 PM positions Vertical Mounted Meters Flow going up works well under all conditions Flow going down works well exc...

Page 3: ...an inlet is often below atmospheric pressure and when the humidity is near 100 there will be condensation formed at these locations The condensate will allow dirt to stick to the sensor more than in t...

Page 4: ...signal and power conduit ports 4 Safety Label and Product ID tag 5 Main I O wiring terminal block for sensor power RS 485 and 4 20 mA outputs TB1 6 AC power input 85 to 265 VAC 50 60 Hz 1 phase 7 Pow...

Page 5: ...he field can damage the sensor wires and result in a loose connection which results in water leakage or other warrantee and hazardous area safety violations TS Version For transmitter separate version...

Page 6: ...lectronics configuration file is matched with the sensor serial number Power and Signal wires 2x conduit ports Conduit seal for Ex applications Sensor Wire Port NPT Figure AH 4 Examples of Unistrut an...

Page 7: ...enclosure at its mechanical interfaces if a fault current was to flow Both internal and external grounds are available see the wiring diagrams DCN 342038 and 342039 For hazardous gas areas wiring goi...

Page 8: ...out few PSI from a regulator Conformal coating of the circuit boards is standard but this only protects against condensation of trapped water vapor which forms from cooling inside the enclosures cond...

Page 9: ...or wiring Braid reinforced pneumatic hose Hydraulic line hose Corrugated Stainless Steel tubing with compression fitting at each end Gas appliance flex fittings may be long enough and are available at...

Page 10: ...supply and will work between 18 and 30 VDC You may also use an unregulated power supply with 50 to 60 Hz ripple as long as the instantaneous voltage is between 18 and 30 VDC Surge currents during sens...

Page 11: ...have a long pull tab attached to the plug so it is easy to remove the plug and connect the AC wiring with a 1 8 screw driver Figure AH 8 AC power pull tab to be removed Discard the pull tab as the po...

Page 12: ...r and wiring For example with a 250 ohm load at 20 mA the voltage drop will be 5 V on the load resistor 19 V across the 4 20 mA output or AO terminals With higher voltage supplies you have correspondi...

Page 13: ...guration files See below for Windows Vista This takes about five seconds at 9600 baud the only supported USB driver baud rate The character format is 1 start bit 1 stop 8 data and no parity The progra...

Page 14: ...and status LEDs with a biased bus allowing the auto polarity detection to work EasySync s devices have a known issue with the Microsoft Windows Vista operating system Contact EasySync at support easy...

Page 15: ...res Conduit seals directly to the enclosures are still needed to meet the explosion proof ratings Peripherally bonded shielded cable glands work well but a simple cable gland and shield pigtail ground...

Page 16: ...braded shield cable then these wires must be clipped in a ferrite to meet the EMC specifications Figure AH 9 shows and example of this type of wiring Figure AH 9 Clip on Ferrite for I O wires to be C...

Page 17: ...nsor electronics and display board must have its connectors fully seated in the PCBs before you carefully screw down the board using the provided screws RTV works as an adhesive to keep the connector...