360208-F 454FTB, 504FTB, 534FTB
14
Keypad
Remote
Keyboard
Function
P
p
Invoke
P
rogram Mode (requires an Access Code);
Or Skip over data with no entry, during data entry
D
d
<BACKSPACE>
Invoke
D
isplay Mode (no Access Code required);
Or delete 1 character to the left, during data entry
L
l
Invoke
L
og Mode (no Access Code required)
E
<ENTER>
Invoke
E
xtended Utilities Mode (requires an Access Code);
Or accept a data entry, during data entry
C
c
C
lear the displayed/entered value, during data entry;
Or acknowledge an active flow meter event
H
h
H
ome or back out of the menu and Return to Run Mode
For the MFT B-Series 2.0 the Quick Setup Code is 123456 and the Tech Code is 654321.
These two access codes can be changed by the user.
A Quick Look up Card for the B-Series 2.0 is provided with each meter. The Quick
Lookup Card is a handy navigational aid for the B-Series 2.0 menu system. Additional
cards are available upon request from Kurz Instruments. Or use this
link
to access a
printable version which can be cut and sized to fit in standard badge-sized laminating
pouches.
Meter Programming
The MFT-B Series 2.0 meter has a Quick Setup Menu that guides the user to configure
the essential parameters to get the meter operational quickly. The following parameters
are included in the Quick Setup Menu:
Meter Tag Name
Flow Units
Flow Area
Probe Depth (for INSERTION meters)
Analog Output #1 Range
Analog Output #2 Range
Setup Run Mode Display
To invoke the Quick Setup Menu, press
P
and enter the Quick Setup Access Code (the
factory default code is ‘123456’). The meter displays the following screen to indicate
that the 4-20mA output signals and values will be ‘frozen’ while using the Quick Setup
Menu.