background image

Kurz

 Instruments, Inc. 

 

831-646-5911 

2411 Garden Road 

 

www.kurzinstruments.com 

Monterey, CA 93940 
 

360208-F    454FTB, 504FTB, 534FTB 

 

 

 

Start Up Guide 
Models 454FTB, 504FTB, and 534FTB. 

August  2010 
 

For MFT B-Series Firmware 2.x 

 

 

 

 

           

 

 

 

 

 

 

 

  

 

Summary of Contents for 454FTB

Page 1: ...struments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 1 Start Up Guide Models 454FTB 504FTB and 534FTB August 2010 For MFT B Series Firmwa...

Page 2: ...et You will need to remove the protective shipping covers from the sensor or flow body before installing the unit A thermal meter must have direct contact with the process fluid to work The meter will...

Page 3: ...depth is adequate to get into the center line where the flow profiles are the most stable The distance from profile disruptions needs to be adhered to for the best repeatability accuracy see the guid...

Page 4: ...Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 4 X Vavg Vs CF v CF v Vavg Vs 0 5 CF v 1 0 Kurz 4xx 454 etc D A ft2 or m2 A D2 4 d L L ft m d in SBCF A A 12dL 5 D CF v v x x x x 0 7 1 0 Vs 2 X 40D X...

Page 5: ...s are used to determine the flow area of the meter An area wizard in the meter setup menu walks you through entering the data and the meter will calculate the flow area 4 The probe will block off some...

Page 6: ...646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 6 X Kurz 5xx 504 etc 5 D Kurz 534 0 D 0 D D X 40D X 20D X 15D 2 Less than two line size changes Less t...

Page 7: ...tially calibrated using long straight runs with short close coupled pipe flanges that minimize the line size discontinuity In many applications field calibration data can be used to reduce the upstrea...

Page 8: ...Kurz Instruments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 8...

Page 9: ...Kurz Instruments Inc 831 646 5911 2411 Garden Road www kurzinstruments com Monterey CA 93940 360208 F 454FTB 504FTB 534FTB 9...

Page 10: ...F 454FTB 504FTB 534FTB 10 Customer I O Panel AC Power 2 2 or 4 4 20 mA AO FLOW Temp 24 VDC Digital I O 454 Vdata Area CF v Customer I O Panel AC Power 2 2 4 20 mA velocity 24 VDC 454 Vdata 4 20 mA AO...

Page 11: ...Series assume the PLC is connected to a loop powered device Potentially Hazardous Area Use and EMI Compliance Please consult the full manual on the provided CD for safety and EMI related issues Sensor...

Page 12: ...age for the meter s AO interface 2 You can self power the 4 20 mA circuit from the flow meter where Kurz provides the 24 VDC excitation voltage The output is not isolated in this case Here your 4 20 m...

Page 13: ...long enough if not twisted up too much to rotate into any of the four positions Navigating Menus The MFT B flow meter with 2 0 and later firmware has a new menu system that is different than the 1 x...

Page 14: ...e changed by the user A Quick Look up Card for the B Series 2 0 is provided with each meter The Quick Lookup Card is a handy navigational aid for the B Series 2 0 menu system Additional cards are avai...

Page 15: ...TAG NAME KURZ MFT B The TAG NAME is entered using the numeric keys on the keypad 0 9 similar to entering a text message on a cellular phone Alphabetic characters and special characters e g etc are ma...

Page 16: ...its for Velocity and Temperature Velocity SFPM SMPS NMPS Temperature DEGF DEGC The duct pipe area blocked off by the insertion of the sensor probe accelerates the flow reading This is adjusted for wit...

Page 17: ...sions are inside dimensions of the duct DUCT WIDTH 36 459 IN DUCT HEIGHT 40 989 IN Use the numeric keys to enter the value in inches or millimeters depending on the flow units specified previously Pre...

Page 18: ...ction displayed In this example FLOW RATE is selected for the Channel 1 AO so the following example screens will prompt for the low and high range of the flow rate being measured The next prompt is fo...

Page 19: ...IC format is selected for the Run Mode Display the meter will prompt for the process variables to display STATIC VARS FLOW TEMP v Use the or v keys to select the desired display option and press E to...

Page 20: ...LOW VEL FLOW TOT VEL FLOW TEMP FLOW TEMP VEL For the Run Mode scrolled format each screen will display the value of only one process variable the description will be on line 1 and the value with units...

Page 21: ...d SAVING CONFIG DATA TO EEPROM Data saved to EEPROM is persistent between power cycles of the flow meter Meter Programming with the USB Port The Quick Setup can also be performed using a PC laptop run...

Page 22: ...and the cursor position is advanced one position to the right for the next character The period and hyphen characters can also be entered in a similar fashion When text input is done press ENTER at th...

Page 23: ...flow while monitoring the indicated flow from the meter This is best left to experienced users or professionals who know how to do it See the manual for more information on Field Calibration The 2 xx...

Page 24: ...point protocol that can be used on the RS 485 port Both the Modbus ASCII or RTU protocols are supported The device address may be any number between 1 and 247 Modbus TCP IP requires a serial to Ether...

Page 25: ...rs and commands field calibration correction factors up to 8 velocity points zero mid span drift check diagnostics alarms sensor calibration purge control damping STP conditions totalizer reset and la...

Reviews: