Kurt J. Lesker Vacuum Mart KJLC 354 Series Troubleshooting Manual Download Page 2

  

Ion gauge cannot be turned 

on 

1.  Pressure exceeds 1.00 x 

10-3 Torr at 4 mA 

emission 

2.  Pressure exceeds 5.0 x 

10-2 Torr at  0.10 mA 

(100 µA) emission  

3.  Emission Control not 

functioning 

1.  Decrease pressure below required value 
2.  Decrease pressure below required value 
3.  Switch to the other filament  

1.  Front Panel Control - page 25 (SETUP IG --> 

FILAMENT NUM --> FILAMENT 1/2) 

2.  RS485 - page 32 

4.  Ion gauge sensor is possibly contaminated.  

1.  Go into R&D Mode 

1.  MENU --> SETUP DISP --> MODE and select 

RESEARCH.  After going to top of menu, it 

should display several things including FVI. 

The first digits are filament voltage, the 

second digits are current.  

i. 

If voltage > 1.7 V and current is > 2.5 A 

at 100 uA setting, there is possible 

contamination 

ii. 

If voltage > 2.3 V and current is > 2.7 A 

at 4 mA setting, there is possible 

contamination. 

2.  Degas sensor if values exceed limits from point 4 

1.  Degassing basic information, page 16 of 

manual 

2.  Front Panel Control --> DEGAS ON 
3.  RS-485 - page 32 of manual 

3.  Re-check R&D Mode Values. If nothing changed, 

sensor is contaminated.  

4.  Contact [email protected] with R&D values 

1.  Possibly need to replace the sensor 

Research Screen shows 

filament Voltage is present 

but filament current stays 

at 0 

1.  Filament is open 

1.  Switch to second filament 

1.  Front panel - page 25 (SETUP IG --> FILAMENT 

NUM --> FILAMENT 1/2) 

2.  RS485 - page 32 of manual 

2.  Check the resistance between filaments 1 and 2 (page 37) 

1.  Remove the ion gauge transducer from the control 

module (page 78, steps 1-5) 

2.  Measure the resistance using an ohmmeter; a good 

intact filament will have a resistance of 0.2 ohms 

(page 73 of manual  

Summary of Contents for Vacuum Mart KJLC 354 Series

Page 1: ...0 of manual 2 Apply gas sensitivity correction factor if applicable or select the appropriate gas from the 354 display menu page 29 of manual 1 Front Panel Control SETUP UNIT GAS TYPE 3 Ion gauge sens...

Page 2: ...s possible contamination ii If voltage 2 3 V and current is 2 7 A at 4 mA setting there is possible contamination 2 Degas sensor if values exceed limits from point 4 1 Degassing basic information page...

Page 3: ...tage 2 3 V and current is 2 7 A at 4 mA setting there is possible contamination 2 Degas sensor if values exceed limits from point 1 a Degassing basic information page 16 of manual b Front Panel Contro...

Page 4: ...nt is 2 7 A at 4 mA setting there is possible contamination 3 Switch to 4 mA emission current and attempt repeated filament starts to clean filament 1 Front panel page 24 EMISSION SEL 4mA 2 RS485 page...

Page 5: ...nsor if values exceed limits from point 1 1 Degassing basic information page 16 of manual 2 Front Panel Control DEGAS ON 3 RS 485 page 32 of manual 4 Re check R D Mode Values If nothing changed sensor...

Page 6: ...ay several things including FVI The first digits are filament voltage the second digits are current i If voltage 1 7 V and current is 2 5 A at 100 uA setting there is possible contamination ii If volt...

Page 7: ...h emission current setting from initial setting and re start the IG make several attempts to establish emission current at both 100 A and 4 mA 1 Front panel page 24 EMISSION SEL 4mA 100 uA 2 RS485 pag...

Page 8: ...e resistance between filaments 1 and 2 page 37 1 Remove the ion gauge transducer from the control module page 42 steps 1 5 2 Measure the resistance using an ohmmeter a good intact filament will have a...

Reviews: