Kurt J. Lesker Vacuum Mart KJLC 354 Series Troubleshooting Manual Download Page 1

 

 

Troubleshooting - Operation Contamination 

KJLC 354 Ion Gauge 

Part Number: KJLC354401xx

 

 

Issue 

Possible Cause 

Possible Solution 

Clear ERROR message 

1.  Must be done before the 

ion gauge can be turned 

on after receiving an 

ERROR message 

1.  SETUP UNIT --> CLR IG ERROR (if using the FRONT 

PANEL CONTROL) 

Display is off / blank 

1.  No power 
2.  The connector may not 

be wired correctly 

  

1.  Check cable connections 
2.  Check cable connections 

1.  Page 12 of manual for the analog pin-out 
2.  Page 13 of manual for the RS485 pin-out 

Readings are very different 

from expected pressure 

1.  Sensor not installed in the 

proper location 

2.  The process gas is 

different from the gas 

used to calibrate the 

gauge (nitrogen) 

3.  Sensor has been dropped 
4.  Gauge sensor is 

contaminated 

5.  Leak in the vacuum 

system 

1.  Ensure sensor is located in the right place (page 10 of 

manual) 

2.  Apply gas sensitivity correction factor if applicable or 

select the appropriate gas from the 354 display menu 

(page 29 of manual) 

1.  Front Panel Control - SETUP UNIT --> GAS TYPE 

3.  Ion gauge sensor is possibly contaminated.  

1.  Go into R&D Mode 

1.  MENU --> SETUP DISP --> MODE and select 

RESEARCH.  After going to top of menu, it 

should display several things including FVI. 

The first digits are filament voltage, the 

second digits are current.  

i. 

If voltage > 1.7 V and current is > 2.5 A 

at 100 uA setting, there is possible 

contamination 

ii. 

If voltage > 2.3 V and current is > 2.7 A 

at 4 mA setting, there is possible 

contamination. 

2.  Degas sensor if values exceed limits from point 3 

1.  Degassing basic information, page 16 of 

manual 

2.  Front Panel Control --> DEGAS ON 
3.  RS-485 - page 32 of manual 

3.  Re-check R&D Mode Values. If nothing changed, 

sensor is contaminated.  

4.  Contact [email protected] with R&D values 

1.  Possibly need to replace the sensor 

4.  Re-check for a leak in the system; metal seals should be 

used when operating below 1x10-7 Torr 

Summary of Contents for Vacuum Mart KJLC 354 Series

Page 1: ...0 of manual 2 Apply gas sensitivity correction factor if applicable or select the appropriate gas from the 354 display menu page 29 of manual 1 Front Panel Control SETUP UNIT GAS TYPE 3 Ion gauge sens...

Page 2: ...s possible contamination ii If voltage 2 3 V and current is 2 7 A at 4 mA setting there is possible contamination 2 Degas sensor if values exceed limits from point 4 1 Degassing basic information page...

Page 3: ...tage 2 3 V and current is 2 7 A at 4 mA setting there is possible contamination 2 Degas sensor if values exceed limits from point 1 a Degassing basic information page 16 of manual b Front Panel Contro...

Page 4: ...nt is 2 7 A at 4 mA setting there is possible contamination 3 Switch to 4 mA emission current and attempt repeated filament starts to clean filament 1 Front panel page 24 EMISSION SEL 4mA 2 RS485 page...

Page 5: ...nsor if values exceed limits from point 1 1 Degassing basic information page 16 of manual 2 Front Panel Control DEGAS ON 3 RS 485 page 32 of manual 4 Re check R D Mode Values If nothing changed sensor...

Page 6: ...ay several things including FVI The first digits are filament voltage the second digits are current i If voltage 1 7 V and current is 2 5 A at 100 uA setting there is possible contamination ii If volt...

Page 7: ...h emission current setting from initial setting and re start the IG make several attempts to establish emission current at both 100 A and 4 mA 1 Front panel page 24 EMISSION SEL 4mA 100 uA 2 RS485 pag...

Page 8: ...e resistance between filaments 1 and 2 page 37 1 Remove the ion gauge transducer from the control module page 42 steps 1 5 2 Measure the resistance using an ohmmeter a good intact filament will have a...

Reviews: