6 Troubleshooting
46
QuickGene-810 Operation Manual
Rev. 1.7
6.3 Error Messages
When an error message appears on the LCD, perform the appropriate action written in the table, and
contact your local representative if necessary.
The error messages displayed on the LCD are as follows:
6.4 When the Buffer is Contaminated
(1) Put DNase/RNase free water to both new conical tubes for wash buffer and
elution buffer, and perform the discharge operation.
For the discharge operation, refer to "
(2) Dispose waste buffer of discharge tray.
(3) Put ethanol to both new conical tubes for wash buffer and elution buffer, and
perform the discharge operation. Dispose waste buffer of discharge tray.
(4) Perform the extraction operation as usual.
Wash buffer or elution buffer
does not flow through the
nozzle.
Is the Pump Tube completely
attached to the joint?
Connect the Pump Tube completely to the
joint.
Is there buffer in the conical
tube?
When the conical tube is empty, put the
buffer into the tube.
Has the tube connecting the
device to the pump become
unfastened from the device?
Contact your local representative.
Phenomenon
Cause
Appropriate action
Message
Contents
Appropriate action
MT ER CR
Carrier motor or air pump has failed.
Contact your local
representative.
The message ("CR",
"HD", "TB", and "WS")
may appear at the same
time, depending on failure
type.
MT ER HD
The nozzle head motor has failed.
MT ER TB
Table motor or lift motor has failed.
MT ER WS
Wash buffer pump or elution buffer pump is not
normal.
POWER ALARM
The power voltage status is not normal.
Contact your local
representative.
EEPROM ER U19
The program in the QuickGene-810 is not
normal.
EEPROM ER U18
EEPROM ER ALL
EEPROM ER CSUM
HISTORY DATA ERR
An error has occurred while writing history
information.
CARRIER POS ERR
Carrier position is not normal.
NOZZLE POS ERR
Nozzle position is not normal.
TABLE POS ERR
Table position is not normal.
LIFT POS ERR
Lift position is not normal.
WD PUMP ERR
Wash buffer pump position is not normal.
CD PUMP ERR
elution buffer pump position is not normal.
PRESSURE ERR
Pressure sensor is not normal.
COVER OPEN
The front cover is open.
Close the front cover after
the necessary operation.
STOP SW PUSHED
The [STOP] switch is pressed, and the
QuickGene-810 operation is stopped.
Press the [RESET] switch.
Summary of Contents for QuickGene-810
Page 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Page 2: ......
Page 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Page 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Page 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Page 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Page 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Page 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Page 76: ......