Kubota 6200 Quick Start Manual Download Page 5

Swinging Bucket Rotor

Tube

Nominal Capacity

24 x 1.5/2mL

16 x 5〜10mL blood collection

tube/15mL

4 x 7〜10mL blood collection

tube/15mL glass

4 x 7〜10mL blood collection

tube/15mL glass

8 x 7〜10mL blood collection

tube/15mL glass

Tube size (mm)
diam. x length

9.5〜11 x 36〜42

12〜17 x 65〜105

12〜17.2 x 86〜110

12〜17.8 x 87〜110

12〜17.2 x 86〜110

Maximum Speed  rpm

4,000

4,000

4,000

 4,000

4,000

Maximum RCF  x

g

2,410

2,410

2,540

 2,500

2,520

Bucket

Code No.

053-4800

053-4930

053-4910

053-4990

053-4900

Adaptor

Code No.

    

055-7400

    

  024-1400 (Rubber cushion)

Tube

Nominal Capacity

4 x 15mL

Conical tube

4 x 15mL

Conical tube

4 x 15mL

Conical tube

8 x 15mL

Conical tube

4 x 50mL

glass

4 x 50mL

Conical tube

4 x 50mL

Conical tube

Tube size (mm)
diam. x length

17 x 121

17 x 121

17 x 121

17 x 121

27〜37 x 95〜110

30 x 117

30 x 117

Maximum Speed  rpm

4,000

4,000

4,000

4,000

4,000

 4,000

4,000

Maximum RCF  x

g

2,580

2,590

2,590

2,580

2,540

 2,610

2,610

Bucket

Code No.

053-4910

053-5010

053-0040

053-4900

053-4990

053-0040

053-5010

Adaptor

Code No.

055-1280

055-1280

*3

*5

*5

*5

*6

*7

*7

*4

*1  Verify plates' strength before centrifugation. Plates in lower position may fracture.
*2  Optional tray (code no.055-6380) is available.
*3  Buckets can be made by request. Please contact us.
*4  5-10mL blood collection tubes which can be used are InsepackⅡ by SEKISUI, Venoject Ⅱ by TERUMO, Neo tube by NIPRO and Vacutainer by BD.
*5  7-10mL blood collection tubes which can be used are InsepackⅡ by SEKISUI, Venoject Ⅱ by TERUMO, Neo tube by NIPRO and Vacutainer by BD.
*6  In case of using glass tubes, rubber cushion (code no.024-1400) is necessary.
*7  Sealing caps and o-rings (sillicon) are equipped with buckets. Autoclavable at 121℃.

SF-240

Max 4,000 rpm

Max 2,610 x

g

SF-240

Max 4,000 rpm

Max 2,610 x

g

Plate

Nominal Capacity

Plate size  mm

Maximum Speed

rpm

Maximum RCF

x

g

Adaptor

Code No.

2 x PCR plate

6 x MTP

2 x DWP

86(W)×128(D)×52(H)

4,000

1,790

2 x PCR plate

12 x MTP

4 x DWP

86(W)×128(D)×90(H)

4,200

3,100

Trays

055-6380

*1

*1

*2

tube-15

tube-15

tube-16

tube-16

tube

1.5mlTube

tube-4

2mlTube

053-4930

BT-65

053-4800

tube-7-3

7ml

tube-7-1

7ml

tube-7-2

7ml

tube-8-2

10ml

tube-8-3

10ml

tube-20

15ml

tube-8-1

10ml

tube-10

15ml

tube-21

50ml

tube-11

50ml

BT-36

055-7400

BT-68

053-4910

BT-68

053-4910

BT-69

053-4990

BT-69

053-4990

BT-67

053-4900

BT-67

053-4900

BT-37

024-1400

053-0040

053-0040

BT-38

055-1280

BT-38

055-1280

053-5010new

053-5010new

tube-6-1

5ml

tube-6-2

5ml

tube-6-3

5ml

tube-7-1

7ml

tube-7-2

7ml

tube-7-3

7ml

tube-8-1

10ml

tube-8-2

10ml

tube-8-3

10ml

tube-20

15ml

6200

PF-206

PF-745S

Tray  

Code No.055-6380

(optional)

 

Plate Rotor

12 x 96 well plate

4 x deep-well plate

2 deep-well + filter 

plate

4 x square-well 

plate

2 x PCR plate

Autoclavable

Ethanol precipitation can be 

carried out at 3,100 xg

Shield covers that can be used 

for biohazard proof are 

equipped with rotor.

Double stack Deep-well plates 

can be used.

Autoclavable

Autoclavable

Flow cytometry

 The number of times allowed for autoclaving is limited. For more details, please contact us.

Rotor has prescribed operating life cycle and years of operation, after that the rotor must be replaced.
For more details, please contact us.

Summary of Contents for 6200

Page 1: ...6200 www centrifuge jp Hybrid High Speed Refrigerated Centrifuge ...

Page 2: ...sity gradient centrifugation also available RCF xg setting operation function and step centrifugal force function can be applied Integrator function can optimize centrifugal effect G sec value integrated RCF value can be measured and the operation with this value is possible Timer can be set up to 990 seconds minutes During the hold condition elapsed running time is displayed Safety quality Equipp...

Page 3: ... 5 2mL 9 5 11 0 x 36 42 15 600 12 x 0 5mL 7 0 7 9 x 25 34 13 330 055 7590 12 x 0 2mL PCR 5 9 6 3 x 20 24 11 320 055 0500 6200 Multi rotor MF 2236A 3 types of buckets can be used with this rotor This helps save cost and time of centrifugation High Speed Centrifugation Maximum speed 15 000rpm Maximum RCF 18 870xg Swinging bucket 24 x 1 5 2 mL 15 000 rpm 18 360xg Fixed angle bucket 24 x 1 5 2 mL 15 0...

Page 4: ...rotective structure for 50mL conical tube caps prevents the tubes from getting broken patented AF 5004CS Biosealed Rotor 4 x 50mL Conical tube Various tubes can be used with adaptors AF 5004CA 4 x 50mL Conical tube AF 5004RA 4 x 50mL 10mL biohazard proof with sealing adaptors code no 055 4280 can be used AF 3806A 6 x 38mL Rotor Name Nominal Capacity Tube size mm diam x length Maximum Speed rpm Max...

Page 5: ... TERUMO Neo tube by NIPRO and Vacutainer by BD 6 In case of using glass tubes rubber cushion code no 024 1400 is necessary 7 Sealing caps and o rings sillicon are equipped with buckets Autoclavable at 121 SF 240 Max 4 000 rpm Max 2 610 xg SF 240 Max 4 000 rpm Max 2 610 xg Plate Nominal Capacity Plate size mm Maximum Speed rpm Maximum RCF xg Adaptor Code No 2 x PCR plate 6 x MTP 2 x DWP 86 W 128 D ...

Page 6: ...No 055 0760 055 1180 tube 1 5mlTube tube 4 2mlTube BT 03 055 4800 BT 07 055 6900 BT 08 055 4740 055 1190 055 0770 tube 6 1 5ml tube 6 2 5ml tube 6 3 5ml tube 7 1 7ml tube 7 2 7ml tube 7 3 7ml tube 9 8mlCPT tube 18 10ml 2 BT 02 053 5840 BT 06 055 0260 1 BT 13 055 4770 055 1200 055 1180 tube 22 1 100ml tube 22 2 100ml tube 12 175ml tube 14 225ml 055 0760 055 6070P 055 1170 tube 11 50ml Table of Roto...

Page 7: ...l 055 1150 055 0750 055 0740 055 4780P tube 10 15ml BT 11 055 4760 tube 21 50ml tube 12 175ml tube 13 200ml tube 14 225ml BT62 1 250ml BT62 2 250ml tube 25 1 tube 25 2 tube 25 3 500mL tube BT 14 055 4850 BT 14 055 4850 055 0780 055 0780 055 1210 n BT 11 055 0430 BT 12 055 0440 BT 12 055 0450 055 0810 055 0800 055 0790 6200 This convenient tube rack can accommodate various sizes such as 5 mL 7 mL 1...

Page 8: ...nt Instruction Service Manual when operating centrifuges Copyright 2020 KUBOTA CORPORATION X SEP Printed in Japan www centrifuge jp 29 9 Hongo 3 chome Bunkyo ku Tokyo 113 0033 Japan Tel 81 3 3815 1331 Fax 81 3 3814 2574 The term of supplying spare parts for repair is 7 years after discontinuation of production except spare parts which we are unable to procure This catalogue is not for distribution...