SECTION 4 – Setup
KTMED Inc. Model OXYPALM
Operator’s Manual
16
[Figure 16]
4-4-3 Pulse rate low alarm limit set
Press
button about 2 seconds, and use
button to display as [Figure 17]. If you want to adjust
Pulse rate low alarm limit, press
. You can set the pulse rate low alarm limit from 30 to 250. (5bpm/
1step)
[Figure 17]
4-5 Time set
Press and hold the menu button
, while pressing and releasing the on /off button
.Display shows as
shown in [Figure 18].
Sequence of time set is year, month, day, hour, and minute.
Change the value with menu button
, press on/off button
when the desired value appears to
advance to the next sequential setting.