6
OPERATION
46
SMARTMAC 400
www.krohne.com
07/2022 - 4008875401 - MA SMARTMAC_400 R01 en
A5.4.34
Stop cleaning
Stop cleaning? (only displayed if Cleaning Mode is selected in A5.4.1
Options:
•
Yes
•
No
A5.5
Relay 2
Settings for Relay 2
A5.5.1
Mode
Operating mode of the relays
Options:
•
Off
•
Limit switch (band)
•
Limit switch (thres.)
•
Field device status
•
USP<645>
•
Loop 1 error current
•
System alarm
A5.5.3
Invert
Invertion of the relay (only displayed if not Off is selected in A4.5.1)
A5.5.5
Set Variable
Settings of the variable that is relating to the relay (only displayed if Limit
switch (band) or Limit switch (thres.) is selected in A5.5.1)
Options:
•
PV
•
SV
•
TV
•
QV
•
4...20 mA L1
•
4...20 mA L2
A5.5.7
Set low alarm
Settings of the lower switching threshold (only displayed if Limit switch
(band) is selected in A5.5.1)
-Min…Max
A5.5.9
Set low alarm hyst.
Settings of the lower hysteresis value (only displayed if Limit switch (band) is
selected in A5.5.1)
-Min…Max
A5.5.11
Set high alarm
Settings of the upper switching threshold (only displayed if Limit switch
(band) is selected in A5.5.1)
-Min…Max
A5.5.13
Set high alarm hyst.
Settings of the upper hysteresis value (only displayed if Limit switch (band) is
selected in A5.5.1)
-Min…Max
A5.5.15
Set threshold
Settings of the switching threshold (only displayed if Limit switch (thres.) is
selected in A5.5.1)
-Min…Max
A5.5.17
Set hysteresis
Settings of the hysteresis value (only displayed if Limit switch (thres.) is
selected in A5.5.1)
-Min…Max
A5.5.19
Set characteristic
Settings of the direction of excess, on which the relay should switch (only
displayed if Limit switch (thres.) is selected in A5.5.1)
Options:
•
Overrun
•
Underrun
Menu level Designation / function
Settings / descriptions