2.9.4 Mounting position
•
Install the flowmeter in horizontal position only in case of the presence of liquids and/or in
steam applications.
•
Horizontal or vertical installation position in case of dry gas.
•
In case of the presence of liquid or steam applications
Figure 2-11: Horizontal or vertical installation
Figure 2-12: Mounting position
WARNING!
Orientate connection box upwards and acoustic path(s) horizontally to avoid liquid in
transducers.