![KROHNE OPTISENS AAC 050 R Installation And Operating Instructions Manual Download Page 20](http://html1.mh-extra.com/html/krohne/optisens-aac-050-r/optisens-aac-050-r_installation-and-operating-instructions-manual_2004174020.webp)
Manual OPTISENS AAC 050
OPTISENS AAC 050
20
5.3
P / PI controller as pulse-pause controller
Main menu
.
.
Set points and
controller versions
Set points
Set point S1
You can choose
.
1.50 mg/l
Controller settings
different controller
.
P range S1
Pulse-Freq. S1
versions for S1 and S2.
0.20 mg/l
00*100/h
Integral time S1
Pulse-Freq. S2
000 sec.
00*100/h
Set point S2
Direction S1
.
0.50 mg/l
Raise
P range S2
Direction S2
0.20 mg/l
Lower
Integral time S2
.
000 sec.
.
Pulse + Pause
The pulse+pause time
Basic settings
10 s
Must be at least twice
Basic settings
.
Least Pulse
As long as the least
Contr. settings
0.1 s
Pulse time.
For a pulse-pause controller you have to set the following parameters:
1) Set points S1 and S2
S1 refers to relay 1, S2 refers to relay 2.
2) P range and integral action time
Adjust a P range > 0. For a P controller set integral time = 0, for a PI controller set an integral time > 0.
3) Pulse-frequencies for S1 and S2
Both frequencies must be set to 00, otherwise the controller will act as an impulse-frequency controller.
4) The acting direction for S1 and S2
Select „raise“ if the dosage raises the measured value.
Select „lower“ if the dosage lowers the measured value.
5) Pulse + pause time
Define a period of time during which the relay is proportionally to the control deviation ON (pulse) or OFF
(pause), respectively.
6) Least pulse time
Set a minimum pulse time that the relay has to at least remain open to allow the actuator to react.