3

 INSTALLATION 

14 

BM26A-5000

www.krohne.com

01/2021 - 4008339901 - HB BM26A-5000 R01 en

3.1  Storage

Do not keep the device in a vertical position before installation. Put the device on its side.

Store the device in its original packing and in dry and dust-free location.

Storage temperature range: 

Storage temperature range: 

Storage temperature range: 

Storage temperature range:  -70

+80

°

C /  -94

+176

°

F

3.2  Transportation

Lift the device carefully. Make sure that the method used to transport the device is adapted to its 
weight and agrees with national regulations.

3.3  Installation - bypass chamber

3.3.1  General notes

The operating conditions that are applicable for your device are given on the device nameplate. 
For more data on where to find the operating conditions data for your device, refer to 

Nameplates

 on page 12. For general technical data about the device, refer to 

Technical data

 on 

page 24.

Make sure that the installation conditions obey the design constraints related to PED. For more 
data, refer to 

EU Pressure Equipment Directive

 on page 6.

Other installation conditions

Make sure that there is no contamination (dirt etc.) or unwanted objects in the bypass 
chamber.

Install the bypass chamber vertically on the tank.

Make sure that the device is correctly installed. The device nameplate is always at the bottom 
of the device.

Select bolts and gaskets (not supplied) that agree with the pressure rating of the process 
connection and the process pressure.

Make sure that mechanical loadings do not cause damage to the process connections. If 
necessary, put support brackets on the device.

3.3.2  Isolation valves

We recommend that you install isolation valves between the device and the tank. The device can 
then be easily removed from the tank for the cleaning procedure. Drain the device only when it is 
isolated from the tank.

DANGER!

Make sure that the outer surface temperature of the device is less than +60

°

C / +140

°

F. If the 

surface temperature is more than +60

°

C / +140

°

F, use the device with precautions that agree 

with health and safety rules and regulations.

WARNING!

Stay away from the process connections. If you stand on the process connections, you can cause 
damage to the device and the installation.

Summary of Contents for BM26A-5000

Page 1: ...rs OPTIFLEX 7200 Guided Radar OPTIFLEX 8200 Guided Radar OPTIWAVE 7400 24 GHz FMCW Radar BW25 Displacer Type BM26A 5000 BM26A 5000 BM26A 5000 BM26A 5000 Handbook Handbook Handbook Handbook KROHNE 01 2...

Page 2: ...or any part thereof without the prior written authorisation of KROHNE Messtechnik GmbH Subject to change without notice 2 www krohne com 01 2021 4008339901 HB BM26A 5000 R01 en Copyright 2021 by KROHN...

Page 3: ...cription 10 2 3 Nameplates 12 2 3 1 Location of device labels 12 2 3 2 Bypass chamber nameplate 13 3 Installation 14 3 1 Storage 14 3 2 Transportation 14 3 3 Installation bypass chamber 14 3 3 1 Gener...

Page 4: ...7 3 Availability of services 19 7 4 Returning the device to the manufacturer 20 7 4 1 General information 20 7 4 2 Form for copying to accompany a returned device 21 7 5 Disposal 22 8 Technical data...

Page 5: ...Certification The manufacturer certifies successful testing of the product by applying the CE marking For more data about the EU Directives and European Standards related to this device refer to the E...

Page 6: ...lculations include exceptional risks such as earthquakes bad weather fire etc The standard design calculation does not take into account the theoretical coefficient of corrosion The product circulatin...

Page 7: ...ess than 0 05 mm 0 002 in diameter Make sure that the device is correctly sealed Do a leakage check between the device and the tank The external pressure Pext must be equal to atmospheric pressure Pat...

Page 8: ...prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards safety requirements and accident preven...

Page 9: ...ngs must be observed without fail Even partial disregard of this warning can lead to serious health problems and even death There is also the risk of seriously damaging the device or parts of the oper...

Page 10: ...op of the bypass chamber A FMCW non contact radar level transmitter is designed to continuously measure the level A TDR Guided Radar level transmitter is designed to continuously measure the level int...

Page 11: ...r 2 Bypass chamber with OPTIFLEX 7200 C TDR Guided Radar level transmitter 3 Bypass chamber with OPTIFLEX 8200 C TDR Guided Radar level transmitter 4 Bypass chamber with OPTIFWAVE 7400 C 24 GHz FMCW r...

Page 12: ...evice is delivered according to your order Check for the correct supply voltage printed on the nameplate Figure 2 3 Location of device nameplates 1 Bypass chamber nameplate 2 Optional level transmitte...

Page 13: ...number 4 Test pressure PT and test date 5 Process temperature range TS 6 Process pressure range PS 7 Bypass chamber volume 8 PED marking PED G1 III PED G1 IV or PED G2 III This marking agrees with NAM...

Page 14: ...o EU Pressure Equipment Directive on page 6 Other installation conditions Make sure that there is no contamination dirt etc or unwanted objects in the bypass chamber Install the bypass chamber vertica...

Page 15: ...BM26A 5000 www krohne com 01 2021 4008339901 HB BM26A 5000 R01 en 3 4 Level transmitter options The level transmitter OPTIFLEX 7200 OPTIFLEX 8200 OPTIWAVE 7400 or BW25 is attached to the bypass chamb...

Page 16: ...the electrical connection procedure refer to the OPTIFLEX 8200 C F S handbook 4 3 OPTIWAVE 7400 24 GHz FMCW radar level transmitter This device is supplied as an option with the bypass chamber For th...

Page 17: ...isolation valves on the top and bottom process connections Do a leak test Energize the electrical equipment i End of the procedure DANGER Devices with Ex approval options Make sure that the bypass ch...

Page 18: ...his device is supplied as an option with the bypass chamber For the device configuration procedure refer to the OPTIFLEX 7200 C F S D handbook 6 3 2 OPTIFLEX 8200 guided radar level transmitter This d...

Page 19: ...ass chamber Use only KROHNE authorized service staff to repair the device 7 3 Availability of services The manufacturer offers a range of services to support the customer after expiration of the warra...

Page 20: ...the health and safety of the personnel the manufacturer may only handle test and repair returned devices that have been in contact with products without risk to personnel and environment This means th...

Page 21: ...r or Serial No has been operated with the following liquid Because this liquid is hazardous to water toxic caustic flammable We have checked that all cavities in the unit are free from such substances...

Page 22: ...ctrical and Electronic Equipment in the European Union Separate collection of WEEE Waste Electrical and Electronic Equipment in the European Union According to the directive 2012 19 EU the monitoring...

Page 23: ...is installed adjacent to the tank and connected to it by two pipes The process conditions in the bypass chamber are the same as those in the tank The measuring principle is also related to the level...

Page 24: ...ed value For more data about the secondary measured value refer to the handbook of the level transmitter option 1 Design Bypass chamber variants Side side process connections 2 Level transmitter optio...

Page 25: ...alloy 2 4819 INCONEL nickel chromium alloy 625 2 4856 5 Stainless steel 1 4404 316L HASTELLOY C 276 alloy 2 4819 INCONEL nickel chromium alloy 625 2 4856 DN80 NPS 3 88 9 mm chamber DN80 NPS 3 88 9 mm...

Page 26: ...with plug 3 8 3 4 NPT female with plug Without 3 8 3 4 NPT female with plug Flange EN All process connection options 9 Flange ASME All process connection options NPS 1 2 or 1 RJ in Class 2500 welding...

Page 27: ...200 guided radar TDR OPTIWAVE 7400 24GHz FMCW radar and BW25 displacer type level transmitters 2 Other variants are available on request 3 The maximum pressure is related to the type of process connec...

Page 28: ...s for the process connections Figure 8 1 BM26A 5000 PN40 Class 300 side side process connections 1 Top flange DN65 NPS 2 1 2 or DN80 NPS 3 for level transmitter option 2 Optional drain with 3 8 3 4 NP...

Page 29: ...ection female plug 76 mm Flange 165 mm Welded pipe 162 mm 4 NPT or G connection male plug 111 mm Flange 165 mm 5 Flange DN65 NPS 2 1 2 73 03 mm Flange DN80 NPS 3 88 9 mm Dimensions inches a b C C c d...

Page 30: ...flange DN65 NPS 2 1 2 or DN80 NPS 3 for level transmitter option 2 Optional drain with 3 8 3 4 NPT female threaded connection with plug or DN15 50 NPS 1 2 2 flange 3 Process connection flange or welde...

Page 31: ...e NPS 1 1 2 197 5 mm Flange NPS 2 222 5 mm 4 NPT or G connection male plug 111 mm Flange 165 mm 5 Flange DN65 NPS 2 1 2 73 03 mm Flange DN80 NPS 3 88 9 mm Dimensions inches a b C C c d e k r t 7 28 1...

Page 32: ...50 Class 1500 side side process connections 1 Top flange NPS 2 Class 1500 NPS 3 Class 900 or NPS 3 Class 1500 for level transmitter option 2 Optional drain with 3 8 3 4 NPT female threaded connection...

Page 33: ...ge DN50 227 mm Flange NPS 1 1 2 220 mm Flange NPS 2 251 mm 4 NPT connection male plug 111 mm Flange 165 mm 5 Flange DN65 NPS 2 1 2 73 03 mm Flange DN80 NPS 3 88 9 mm Dimensions inches a b C C c d e k...

Page 34: ...Class 2500 or NPS 3 Class 2500 for level transmitter option 2 Optional drain with 3 8 3 4 NPT female threaded connection with plug or DN15 50 NPS 1 2 2 flange 3 Process connection flange or welded pip...

Page 35: ...ange NPS 1 165 mm Flange NPS 2 251 mm 4 NPT connection male plug 111 5 mm Flange 165 mm 5 Flange DN65 NPS 2 1 2 73 03 mm Flange DN80 NPS 3 88 9 mm Dimensions inches a b C C c d e k r t 7 28 1 13 39 11...

Page 36: ...ort bracket Figure 8 5 Support bracket Dimensions mm a b c d e f g h j k m n p s t u Support bracket 73 3 52 28 5 4 8 4 15 15 50 80 26 12 16 5 33 4 58 Table 8 10 Support bracket dimensions in mm Dimen...

Page 37: ...0 20 20 38 50 100 Maximum process pressure barg EN 1092 1 PN40 40 40 40 40 40 40 40 PN100 100 100 100 100 100 100 100 PN160 160 160 160 160 160 160 160 PN250 250 250 250 250 250 250 250 PN400 400 400...

Page 38: ...8 337 1 318 297 1 285 7 274 2 ASME B16 5 Class 150 14 8 13 6 12 10 2 8 4 6 5 Class 300 36 3 33 7 31 8 29 7 28 5 27 4 Class 600 77 1 71 2 66 7 63 1 61 58 9 Class 1500 192 4 178 4 167 158 1 152 2 147 1...

Page 39: ...25 9 3625 9 3625 9 3625 9 3625 9 ISO 228 1 G 580 2 580 2 580 2 580 2 580 2 580 2 580 2 ASME B1 20 1 NPT 1450 4 1450 4 1450 4 1450 4 1450 4 1450 4 1450 4 Table 8 14 Maximum process pressure psig for a...

Page 40: ...S 40 40 40 40 40 40 34 4 40S 100 100 100 100 100 100 86 1 ISO 228 1 G 40 40 40 40 40 40 34 4 ASME B1 20 1 NPT 100 100 100 100 100 100 86 1 Table 8 16 Maximum process pressure barg for a given maximum...

Page 41: ...4 1450 4 1450 4 1450 4 1248 8 ISO 228 1 G 580 2 580 2 580 2 580 2 580 2 580 2 498 9 ASME B1 20 1 NPT 1450 4 1450 4 1450 4 1450 4 1450 4 1450 4 1248 8 Table 8 18 Maximum process pressure psig for a giv...

Page 42: ...d There is one process connection at the top and one more at the bottom of the measuring chamber Options include a level transmitter axial or lateral process connections vents and drains Electromagnet...

Page 43: ...NOTES 10 43 BM26A 5000 www krohne com 01 2021 4008339901 HB BM26A 5000 R01 en...

Page 44: ...ing solutions Engineering commissioning calibration maintenance and training services Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 103...

Reviews: