aeroplus m
Menu and advanced settings
33
10.6 Upper limit for the therapy setting
In order to limit the therapy range, a lower (minimum oxygen volume, O2-Min) and a upper
limit (maximum oxygen volume, O2-Max) can be stored. These limits ensure the operation
only between the applied limits.
Display/ Message
Description
To set the upper limit, select menu item
O2-Max.
in the main menu;
the selection is inverted with white back ground color.
NOTE!
Already in the main menu, the current setting can be controlled
(here: setting 5.0). It is not necessary to call the submenu.
Confirm selected menu item with the menu button
.
To increase the setting press the arrow-up-button
.
To decrease the setting press the arrow-down-button
.
To leave this menu, press the menu button
(=
return
)
.
NOTE!
The upper limit can only be set between the lower therapy limit
and 5.0. If applicable, the lower limit must be modified first; not
till then the upper limit can be adjusted as desired.
NOTE!
If the lower limit has the same value as the upper limit, this will
freeze the therapy setting.