KORAD KEL102 Calibration Instruction Download Page 2

4.

 

Verify

 

DA

 

and

 

AD

 

when

 

the

 

maximum

 

voltage

 

is

 

up

 

to

 

90V:

 

Method:

 

set

 

the

 

power

 

supply

 

to

 

100V

 

and

 

the

 

maximum

 

current

 

output

 

to

 

1A.

 

Set

 

the

 

multimeter

 

to

 

the

 

voltage

 

range.

 

Combine

 

the

 

power

 

supply

 

and

 

the

 

multimeter

 

into

 

the

 

load

 

interface

 

at

 

the

 

same

 

time.

 

Turn

 

on

 

ON/OFF

 

switch

 

and

 

adjust

 

the

 

knob

 

to

 

make

 

the

 

voltage

 

on

 

the

 

multimeter

 

display

 

90V.

 

And

 

press

 

the

 

left

 

and

 

right

 

keys

 

to

 

adjust

 

the

 

increased

 

and

 

reduced

 

value.

 

Then

 

press

 

ENTER

 

to

 

enter

 

the

 

next

 

check,

 

or

 

press

 

ESC

 

to

 

return

 

to

 

the

 

previous

 

check.

 

   

5.

 

Check

 

the

 

zero

 

point

 

of

 

DA

 

low

 

current:

 

Method:

 

set

 

the

 

power

 

supply

 

to

 

5V

 

and

 

the

 

maximum

 

current

 

output

 

to

 

1A.

 

Set

 

the

 

multimeter

 

to

 

the

 

current

 

range.

 

Connect

 

the

 

power

 

supply

 

and

 

the

 

multimeter

 

to

 

the

 

load

 

interface.

 

Turn

 

on

 

the

 

ON/OFF

 

switch

 

and

 

adjust

 

the

 

knob

 

to

 

make

 

the

 

multimeter

 

current

 

display

 

between

 

1mA

 

and

 

2mA.

 

And

 

press

 

the

 

left

 

and

 

right

 

keys

 

to

 

adjust

 

the

 

increased

 

and

 

reduced

 

value.

 

Then

 

press

 

ENTER

 

to

 

enter

 

the

 

next

 

check,

 

or

 

press

 

ESC

 

to

 

return

 

to

 

the

 

previous

 

check.

 

     

6.

 

Check

 

the

 

zero

 

point

 

of

 

DA

 

high

 

current:

 

Method:

 

set

 

the

 

power

 

supply

 

to

 

5V

 

and

 

the

 

maximum

 

current

 

output

 

to

 

1A.

 

Set

 

the

 

multimeter

 

to

 

the

 

current

 

range.

 

Connect

 

the

 

power

 

supply

 

and

 

the

 

multimeter

 

to

 

the

 

load

 

interface.

 

Turn

 

on

 

ON/OFF

 

switch

 

and

 

adjust

 

the

 

knob

 

to

 

make

 

the

 

current

 

on

 

the

 

multimeter

 

display

 

between

 

1mA

 

and

 

2mA.

 

And

 

press

 

the

 

left

 

and

 

right

 

keys

 

to

 

adjust

 

the

 

increased

 

and

 

reduced

 

value.

 

Then

 

press

 

ENTER

 

to

 

enter

 

the

 

next

 

check,

 

or

 

press

 

ESC

 

to

 

return

 

to

 

the

 

previous

 

check.

 

   

7.

 

Verify

 

DA

 

and

 

AD

 

when

 

the

 

maximum

 

value

 

of

 

the

 

current

 

low

 

gear

 

is

 

3A:

 

Method:

 

set

 

the

 

power

 

supply

 

to

 

5V

 

and

 

the

 

maximum

 

current

 

output

 

to

 

5A,

 

set

 

the

 

multimeter

 

to

 

the

 

current

 

range,

 

connect

 

the

 

power

 

supply

 

and

 

the

 

multimeter

 

to

 

the

 

load

 

interface,

 

turn

 

on

 

the

 

ON/OFF

 

switch,

 

and

 

adjust

 

the

 

knob

 

to

 

make

 

the

 

current

 

on

 

the

 

multimeter

 

display

 

3A.

 

And

 

press

 

the

 

left

 

and

 

right

 

keys

 

to

 

adjust

 

the

 

increased

 

and

 

reduced

 

value.

 

Then

 

press

 

ENTER

 

to

 

enter

 

the

 

next

 

check,

 

or

 

press

 

ESC

 

to

 

return

 

to

 

the

 

previous

 

check.

 

 

8.

 

Verify

 

DA

 

and

 

AD

 

when

 

the

 

maximum

 

value

 

of

 

the

 

current

 

low

 

gear

 

is

 

9.9A:

 

Method:

 

set

 

the

 

power

 

supply

 

to

 

5A

 

and

 

the

 

maximum

 

output

 

to

 

11A,

 

set

 

the

 

multimeter

 

to

 

the

 

current

 

range,

 

connect

 

the

 

power

 

supply

 

and

 

the

 

multimeter

 

to

 

the

 

load

 

interface,

 

turn

 

on

 

the

 

ON/OFF

 

switch,

 

and

 

adjust

 

the

 

knob

 

to

 

make

 

the

 

current

 

on

 

the

 

multimeter

 

display

 

9.9A.

 

You

 

can

 

press

 

the

 

left

 

and

 

right

 

keys

 

to

 

increase

 

or

 

decrease.

 

Then

 

press

 

ENTER

 

to

 

save

 

and

 

complete

 

the

 

verification,

 

or

 

press

 

ESC

 

to

 

return

 

to

 

the

 

previous

 

verification.

   