23
typical location in a
typical commercial
or hospital
environment.
Note 1: U
T
is the AC mains voltage prior to application of the test
level.
6.3 Electromagnetic Immunity: For non-life support
equipment and systems
The device is intended for use in the electromagnetic
environment specified below. The customer or the user of the
device can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output
power of the communications equipment.
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity
test
IEC60601
Test level
Compliance
level
Electromagnetic
environment-guidance
Conducted
RF
IEC
61000-4-6
Radiated
3V effective
value
150kH-80MHz
3V/m
3V effective
value
3V/m
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the device including
cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the