![KONICS KRN1000 SERIES User Manual Download Page 19](http://html.mh-extra.com/html/konics/krn1000-series/krn1000-series_user-manual_1994598019.webp)
2
Specifications
© Copyright Reserved KONICS Co., Ltd.
19
※
2.
◎
At room temperature (23
℃
± 5
℃
)
RTD Cu50Ω (-200≤T≤200): (±0.1% F.S. or ±1.5
℃
, select the higher one) ±1-digit
RTD DPt50Ω (-200≤T≤500): (±0.1% F.S. or ±1.5
℃
, select the higher one) ±1-digit
Thermocouple R, S, C, G type
(0≤T≤100): (±0.1% F.S. or ±4.0
℃
, select the higher one)
±1-digit
Thermocouple U, T type (-
100≤T≤400): (±0.1% F.S. or ±2.0
℃
, select the higher one)
±1-digit
Thermocouple B type, below 400
℃
: There is no accuracy standards.
All thermocouples, below -100
℃
: (±0.3% F.S. or ±4.0
℃
, select the higher one) ±1-digit
◎
Out of room temperature range
RTD Cu50Ω (-200≤T≤200): (±0.2% F.S. or ±3.0
℃
, select the higher one) ±1-digit
RTD DPt50Ω (-200≤T≤500): (±0.2% F.S. or ±3.0
℃
, select the higher one) ±1-digit
※
3.
Input/Output is different by option. Refer to ‘1.3 Ordering information’.
※
4.
For supplying power for transmitter, it is recommened to use shield cable to reduce noise.
※
5.
RS422/485, ethernet, USB device communication outputs are not used at the same time.
※
If sensor input line is longer, it is recommended to use shield cable to reduce noise.
Before the wiring, turn OFF the unit.
Be sure that the polarity of power for transmitter. If not, it may cause product damage.
When supplying the over voltage at sensor input, it may cause product damage.
Summary of Contents for KRN1000 SERIES
Page 1: ... Copyright Reserved KONICS Co Ltd i USER MANUAL ...
Page 2: ...ii Copyright Reserved KONICS Co Ltd ...
Page 8: ...Safety Precautions viii Copyright Reserved KONICS Co Ltd ...
Page 16: ...1 Product Introduction 16 Copyright Reserved KONICS Co Ltd ...
Page 22: ...3 Dimensions 22 Copyright Reserved KONICS Co Ltd ...
Page 85: ...9 Troubleshooting Copyright Reserved KONICS Co Ltd 85 ...
Page 86: ...9 Troubleshooting 86 Copyright Reserved KONICS Co Ltd KUM E 0180C_20160923 ...