Customer Support
1.800.366.5343
–
4.7.7 Download the Calibration Files to the
Panel
After closing the Calibration window, you will be
back where you left off, at the iDetector
Calibrate
tab. Now we need to download the
new calibration files to the panel.
1. Within the Calibrate tab, click on the
Gain
tab, then click the
ReadStatus
button.
2. Highlight
Index 1
, then click the
SelectFile
button. This will change the
Index 1 status to enable.
3. Click the
DownloadFile
button, and
then click the
Path
button from the
popup window.
4. Browse to the
C:\opal\data\KGxxxxxxxxxxx\Correct
\Default folder and select the
gain_2304x2800.gn
file and click
Open
.
Note
: if the gain file does not appear,
verify the file type dropdown is set
to
gain files(*.gn)
.
5. Type
1
in the
FileIndex
field and then
click
OK
.
6. After the gain file has been downloaded
to the panel. Click
OK
.