background image

 

  1 

 

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ViZion DR+ 
1417V2 
Wireless 
Panel 

Dealer 
Installation 
Manual 

Summary of Contents for ViZion DR+ 1417V2

Page 1: ...1 ViZion DR 1417V2 Wireless Panel Dealer Installation Manual ...

Page 2: ... Synchronization Ethernet and Power Cables 7 1 6 ViZion DR Wireless Panel Battery Charger 8 1 7 Additional Components 9 2 Hardware Installation 9 2 1 Fixed Room Hardware Installation 9 2 2 Mobile Hardware Installation 10 3 Network Installation 11 3 1 Fixed Room Network Configuration 11 3 2 Mobile Network Configuration 12 Software Installation and Configuration 15 4 Software Installation 15 4 1 Ins...

Page 3: ...ort konicaminolta com Revision History REV DATE DOCUMENT NAME REASON FOR CHANGE 1 0 6 27 2018 ViZion DR 1417V2 Wireless Panel Installation Manual New Manual iRay SDK Version 2 and Ultra version 4 2 9 0 Distributed by Konica Minolta Healthcare Americas Inc 2217 US Highway 70 East Garner NC 27529 1 800 366 5343 ...

Page 4: ...ll configure and calibrate the ViZion DR 1417V2 Wireless Panel For more detailed information on the operation and specifications of your wireless detector please review the following documentation Manual References 059 201 04 Mars1417V2 User Manual A0 pdf Only qualified service personnel should attempt to install modify service or operate the ViZion DR Wireless Panel detector ...

Page 5: ...your ViZion DR system 1 1 Passive Installation ViZion DR 1417V2 Wireless Panel Detector ASUS Dual Band Wi Fi Gigabit AP 75 ft Cat5 Ethernet Cable Acquisition PC Desktop or Laptop Windows 10 1 2 Integrated Installation The following additional equipment is required for integration with a generator G I P S Box Power Supply Synchronization Cable 50 ft Female to Female Serial Cable Tripp Lite USB to S...

Page 6: ... E Infrared window Infrared device sensor F Antenna Wi Fi antennas three G Reserved H Maintenance cover For service engineer to maintenance I Battery lock The lock button for detaching battery J Detector label Serial number and product information 1 3 2 LED Indicators 1 Battery Indicator Green battery charge is 10 Orange battery charge is 10 2 Link Indicator Blue solid wireless connection is conne...

Page 7: ...erial port male 5 Panel Power Panel 1 and Panel 2 1 5 Synchronization Ethernet and Power Cables Your ViZion DR wireless detector includes the following accessories 75 ft Cat5 Ethernet Cable 50 ft Female to Female Serial Cable Tripp Lite USB to Serial Adapter Power Supply Synchronization Cable 1 5 1 Synchronization Cable Review the available signal wires in the following figures prior to connecting...

Page 8: ...een light displays when power is on D Hand pull position E Limit ball plug F DC Jack 24V DC input 1 6 1 Battery Charger Capacity Indicators 1 6 2 Connector Ports It is alternatively possible to trickle charge the battery inside the panel by connecting the DC power cable from the DC medical adapter power supply to the DC input of the detector To preconfigure the panel for alternate wireless connect...

Page 9: ...ve all required hardware please consult the equipment list which can be located at the beginning of Section 1 The following sections provide step by step instructions to complete the hardware installation The connections between the various devices are more or less straightforward however only a qualified service technician should attempt to connect the synchronization cable to the generator and b...

Page 10: ...ucky interface Note You can find detailed information on connecting your cable to a specific generator in the appropriate appendix in the Engineering folder located on the desktop of the Ultra workstation 2 1 2 Connecting the G I P S Box to the Acquisition PC To connect the G I P S box to the acquisition PC for integration 1 Connect the 50 F to F serial cable to Port 4 on the G I P S box 2 Connect...

Page 11: ...n PC and ASUS access point all come preconfigured These instructions show how to configure the devices as needed The goal is to ensure that each device can communicate with one another which requires that they are configured on the same network segment The following network equipment is required for fixed room environments ASUS Dual Band Wi Fi Gigabit AP 75 Cat5 Ethernet Cable Link Cable to precon...

Page 12: ...to Client mode to connect the laptop to an external access point device that the panel will connect to recommended method or put in AP mode to allow the panel to broadcast a wireless network for the laptop to connect to alternative method 3 2 1 ViZion DR 1417V2 Wireless Panel in Client Mode Recommended method When the panel is in Client mode the panel will connect automatically to the external TP ...

Page 13: ...anges 6 Connect the ethernet RJ 45 end of the Link cable to the LAN Ethernet port on the laptop computer 7 Connect the mini HDMI end of the Link cable to the Link port on the side of the panel Note follow the instructions below AFTER COMPLETING ALL STEPS in this installation manual For now leave the Link cable connected until installation is complete Unpack the TP Link AC750 AP Make sure the mode ...

Page 14: ...hernet port the same as in Section 3 2 1 Once the installation is complete follow the steps below to configure the Wireless network adapter for use in wireless AP mode 1 From Control Panel select Network Connections and right click on the Intel Wireless adapter and select Properties 2 Select Internet Protocol Version 4 TCP IPv4 and click Properties 3 Select Use the following IP address radio butto...

Page 15: ... C Redistributable for Visual Studio 2013 before installation of the Ultra software and latest venu driver packs 1 Download the 32 bit and the 64 bit versions of the redistributable files You will need to install both versions 2 Start with installing the 32 bit executable file x86 3 Check the box I agree to the license terms and conditions and click Install 4 The installation usually takes less th...

Page 16: ...er file to the desktop of the acquisition PC 2 Right click the file and select Run as Administrator 3 Click Yes when prompted for Do you want to backup 4 When the installation has completed click Close 4 3 Configuration of Ultra Software 1 Double click on the Ultra icon to launch the software 2 If this is a new installation you will be presented with a prompt informing you that this is the first t...

Page 17: ...rative box The box will change to green when a valid password has been entered 3 After entering the password click Apply to activate all the all administrative configuration options 4 3 2 Create Technologists Ultra Only For DR technician support purposes add a new technologist for Konica Minolta KM If you know the site technologist s names you can also add those at this time 1 Click on the Tools A...

Page 18: ...M2 panels should be configured to use First Generation Venu drivers ViZion DR 1417V and 1717M2 5 panels should be configured to use Second Generation Venu drivers ViZion DR 1417V2 and 1012V2 panels should be configured to use Third Generation Venu drivers 1 Before installing the new venu driver pack double click on the Konica Minolta logo located in the bottom left corner of the Options menu This ...

Page 19: ...nch button 7 This will launch the Device Manager window Click the Shutdown Devices button 8 Wait until the Devices status changes from Initialized to Shutdown 9 After shutting down devices check the boxes for Download Driver and Install Driver next to Venu DR in the middle of the Device Manager window 10 Click the Perform Driver Actions button and it will proceed to download the venu driver pack 1...

Page 20: ...o complete 16 Once completed you should verify that Venu installed Note if it shows Venu failed instead of Venu Installed the download did not fully transfer or got corrupted during download Check to make sure Windows Firewall is disabled and AV is excluding the C opal folder Then rerun the driver download and install again 17 Click Initialize Devices and exit the Device Manager window by clicking...

Page 21: ...ration Pool section 3 Click on ViZion DR and the panel device configuration section will appear below it 4 Scroll down to the Device section and click the button with 3 dots to the right of S N field This will perform a scan and automatically populate the S N and Path fields 5 By default Check Shock Log is unchecked and Manage Automatically is checked 6 Select the drop down next to Parameters Temp...

Page 22: ...the Detector tab confirm that Trigger mode is set to TriggerMode_FreeSync If it is already set skip to the next section for Wi Fi configuration of the panel 4 If the panel was in a different mode change it to TriggerMode_FreeSync 5 Click Write and wait for the State at the bottom to change back to Ready highlighted in green 6 Click Write ROM and wait for the State at the bottom to change back to R...

Page 23: ... Wi Fi Configuration in AP Mode Alternative connection method 1 Within the Detector tab click on the Wifi tab and select AP mode 2 Click Add button and configure the fields with the following info SSID enter the serial number for the panel It is printed on the bar code label on the back of the panel Key 12345678 Security Wifi_WPA_WPA2_PSK Frequency Wifi_Freq_5GHz Country Wifi_Country_US Band Wifi_...

Page 24: ...able connected to the detector 3 Select the Calibrate tab check the radio button for PostOffset and click the Start Generate Templates button The Calibration wizard will open into a new window presenting the Welcome screen 4 Set X ray environment Tube SID 50 or higher at least greater than 47 No Collimation open collimator blades at least 1 past the edges of the panel on all sides No Grid remove a...

Page 25: ... 5 blank gain exposure frame icons Each exposure frame is numbered sequentially with the kV to use noted next to each one A green box surrounds the current exposure frame 8 Set your techniques to 70kV 4 7 mAs and take your first exposure Note after taking an exposure the current gain exposure frame will still be surrounded in green It will either have a green checkmark if it was successful or a ye...

Page 26: ...Calibration screen requires 19 exposures The left hand side of the screen tells you the suggested kV and mAs to use for the current exposure The right hand side of the screen is divided into 19 blank defect exposure frame icons Each exposure frame is numbered sequentially with the kV to use noted next to each one A green box surrounds the current exposure frame 13 Set your techniques to 70kV 1 1 m...

Page 27: ... the Prep button and take another exposure 15 If the exposure was acceptable message in green click the Next button to move to the next frame 16 Set your techniques to 40kV 4mAs and take your second exposure 17 If the exposure was acceptable click Next to move to the next frame 18 Set your techniques to 120kV 1 2mAs and take your third exposure 19 If the exposure was acceptable click Next to move ...

Page 28: ...tinue to take exposures with techniques set to 70kV 7mAs for the remaining frames 5 19 23 After taking all 19 defect exposures click the Back button if you want to retake any previous exposures then click Prep and retake the exposure Otherwise if you are satisfied with all exposures click the Next button in the top right corner 24 Wait for the Defect Map Generation to complete and then click Next ...

Page 29: ...ate tab click on the Gain tab then click the ReadStatus button 2 Highlight Index 1 then click the SelectFile button This will change the Index 1 status to enable 3 Click the DownloadFile button and then click the Path button from the popup window 4 Browse to the C opal data KGxxxxxxxxxxx Correct Default folder and select the gain_2304x2800 gn file and click Open Note if the gain file does not appe...

Page 30: ...ton This will change the Index 1 status to enable 9 Click the DownloadFile button and then click the Path button from the popup window 10 Browse to the C opal data KGxxxxxxxxxxx Correct Default folder and select the defect_2304x2800 gn file Click Open Note if the defect file does not appear verify the file type dropdown is set to defect files dft 11 Type 1 in the FileIndex field and then click OK ...

Page 31: ...rder 14 Check the radio buttons for HWPostOffset HWGain and HWDefect 4 7 9 Closing iDetector It s important to close iDetector properly in order to sync the changes made in iDetector with Ultra UAI 15 Select the Home tab and click the Close button 16 Select Override Config 17 Disconnect the Link cable from both ends and store it in a secure place for future reconfiguration purposes Mobile techs sh...

Page 32: ...ual Ultra 4 2 9 0 32 Customer Support 1 800 366 5343 HCITsupport konicaminolta com Note if you configured the panel for Client mode go back to Section 3 2 2 and follow the instructions for configuring the wireless adapter and installing the TP Link AP ...

Reviews: