30
5
Change the CCF using the [
▲
] and [
▼
] keys.
Each time the [
▲
] or [
▼
] key is pressed, the value will change in 0.001
steps (settable range: 0.500 to 2.000).
Holding the [
▲
] or [
▼
] key will cause the
displayed value to change continuously.
z
To exit setting, press the [CALL] key.
6
When you have changed the value, press the [SET] key.
The CCF will be set.
z
To check the currently set CCF, press the [CALL] key. The currently set CCF will be displayed while the
[CALL] key is held down.
k
Entering the reference value
If you have a strictly controlled reference light source, the correction factor (CCF) can be
obtained using the CCF function.
If the reference illuminance value of the target light source is known, just enter the reference
value to the instrument. The CCF will be calculated and set automatically.
■
Entering the Reference Value
1
Press the [NORM] key to activate normal mode.
z
The reference value can be set in normal mode only.
2
Push in the hold button (HOLD).
3
Open the slide cover and press the [
CCF
] key.
CCF mode will be activated.