387
10 .2 Error codes and countermeasures
Chapter 10
Error code
Displayed message
Problems/countermeasures
C27600 -
C27604
Initialization Error
Check the destination connected to .
C27605
Health check error
C27606 -
C27608
Scheduled check error
C2761*
Error in study completed notification to RIS
C27620 -
C27626
Patient search error
C27628
C27629
C2763*
C2764*
Folder check error
C2770*
C2773*
Communication error occurred in RIS-OUT communi-
cation .
Check the communication destination .
Check the network status .
C2774*
Communication error occurred in RIS-IN communica-
tion .
Check the communication destination .
10 .2 .2 Software errors
Error codes starting with "C3" through "C7" are indicating software errors .
Error code
Displayed message
Problems/countermeasures
C35500
Failed to initialize .
Restart this device .
C35501
There was an error with reading the setting file.
Check the setting file and update it.
C35502
A time-out occurred .
MWM search request response waiting time-out oc-
curred .
Check the connection destination .
C35503
A socket error occurred .
An error occurred in communication with DICOM .
(1) Check the destination connected to, the network
status, and connection settings .
(2) Restart the device .
C35504
Channel mismatch occurred
Contact Konica Minolta technical representatives .
C35520
Failed to initialize .
Restart this device .
C35521
There was an error with reading the setting file.
Check the setting file and update it.
C35522
A time-out occurred .
PPS result notification time-out occurred.
Check the connection destination .
C35540
Failed to initialize .
Restart this device .
C35541
There was an error with reading the setting file.
Check the setting file and update it.
C35542
A time-out occurred .
Detach result notification response timeout occurred.
Check the connection destination .
C35543
Channel mismatch occurred
Contact Konica Minolta technical representatives .
C35560
Failed to initialize .
C35561
There was an error with reading the setting file.
C35562
A time-out occurred .
C35563
Socket error has occurred .
Check the network equipment and communication
cable connection status .
C35564
Exceeded the maximum number of strings obtained.
Contact Konica Minolta technical representatives .
C36020
At PrintComposer process, execution error occurred.
(1) Try performing the operation again .
(2) Contact Konica Minolta technical representatives .
C36021
At PrimtComposer process, Fatal error occurred . Con-
tact a service representative .
Contact Konica Minolta technical representatives .
Continued on the next page
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...