1 .2 Safety precautions
WARNING
• Be sure to implement the following cyber-security
measures:
– Provide for an adequate level of information se-
curity measures within the facility using this de-
vice, and other facilities connected to the facility
via networks .
– Isolate the device from the connected network
immediately upon infection by malware, concur-
rently prohibiting the use of the infected device .
– Upon infection of the device by malware, or
the potential infection thereof, promptly contact
Konica Minolta service representative .
• Manage the account information (user name, ID,
etc .), password and other sensitive information as
follows:
– Use a combination of upper- and lower-case al-
phabet and numeric characters to prevent easy
identification by malicious parties.
– Do not leave records of sensitive information to
avoid uncertified disclosure.
– Passwords should consist of at least eight char-
acters .
–
Passwords should be assigned with expiration
dates, and changed on a regular basis .
1 .2 .8 Precautions for each
function
WARNING
Precautions on stitching combination processing
• Adjust the combination position by using the mark-
ers as supplementary standards in addition to body
parts of the subject . It is also recommended that
you arrange the measure in the combined image
part during exposure to simplify the confirmation of
the combined parts .
Precautions for One Shot stitching exposure
•
One Shot stitching exposure outputs the combined
picture by correcting the images' overlapping part
captured by DR panel . Depending on the capturing
conditions, lateral stripes or decrease in contrast
might occur in a correction area, or decrease in
contrast might also occur if a structure horizontally
long overlaps with the correction area . Make sure
the correction area and the parts in an examination
are not overlapping .
• Make sure to observe the number of DR panels
that are loaded in the stitching holder through the
exposure screen before conducting the exposure.
Precautions for image measurements
•
Sometimes the exposure orientation causes ac
-
cidental errors with regard to the actual length of
the subject . Use while taking into account any error
between image and measured results .
Precautions when running HS processing
• The noise is reduced by HS processing, and you can
expect signals that are unclear due to noise to become
clearer, but signals that do not exist will not be visible.
• When reviewing whether to reduce dosage by means
of HS processing, consider lowering the signal com-
ponent by means of reducing dosage and always
check whether images are suitable for diagnosis .
Summary of Contents for CS-7
Page 1: ...0197 Operation Manual EN 15 DIRECT DIGITIZER Version 1 30 ...
Page 2: ......
Page 8: ...8 ...
Page 9: ...9 Introduction ...
Page 18: ...18 ...
Page 30: ...30 ...
Page 31: ...31 Chapter 2 Product Overview This chapter describes the overview of this device ...
Page 34: ...34 ...
Page 74: ...74 ...
Page 84: ...84 ...
Page 120: ...120 ...
Page 197: ...197 Chapter 7 Various Functions This chapter describes the functions of this device ...
Page 383: ...383 Chapter 10 Error Codes This chapter describes the error codes and countermeasures ...
Page 413: ...413 Chapter 11 Maintenance This chapter describes the items that require periodic maintenance ...
Page 416: ...416 ...
Page 417: ...417 Chapter 12 Specifications This chapter describes the specifications of this device ...
Page 422: ...422 ...
Page 423: ......
Page 424: ...A47FBA01EN15 2017 01 12 JD ...