Certificate No : JQA-E-80027
Registration Date : March 12, 1997
Certificate No : LRQ 0960094/A
Registration Date : March 3, 1995
For correct use and for your safety, be sure to read the instruction
manual before using the instrument.
•
Be sure to use the specified power supply voltage. Improper
connection may cause a fire or electric shock.
SAFETY PRECAUTIONS
9242-4810-20 BHDPK
Printed in Japan
• KONICA MINOLTA, the Konica Minolta logo and symbol
mark, and "Giving Shape to ideas" are registered
trademarks or trademarks of KONICA MINOLTA, INC.
• Displays shown are for illustration purpose only.
• The specifications and appearance shown
herein are subject to change without notice.
• Other company names and product names
used herein are trademarks or registered
trademarks of their respective companies.
©2015-2017 KONICA MINOLTA, INC.
Main Specifications
Dimensions (Units:mm)
System Diagram
Model
CS-150
CS-160
Measuring angle
1°
1/3°
Optical system
SLR viewing system, f = 85 mm F2.8
Angle of view
9° (with diopter adjustment)
Relative spectral
responsivity
Closely matches CIE 1931 color matching function (x (
λ
), y (
λ
),
z (
λ
))
Minimum measuring
area(diameter)
14.4 mm
(1.3 mm when close-up lens is
used)
4.5 mm
(0.4 mm when close-up lens is
used)
Minimum measuring
distance (From the
measuring distance
reference plane)
1,012 mm
(213 mm when close-up lens is used)
Color notations
(Absolute value) L
v
, x, y (Y, x, y), L
v
, u', v', L
v
, T
cp
, duv, XYZ,
L
v
,
λ
d
, P
e
Measurement mode
(Luminance) Instantaneous value, maximum/minimum
value, luminance difference (
Δ
)/luminance
ratio (%)
(Chromaticity) Instantaneous value, chromaticity difference
(
Δ
)
Measurement time
Auto: 0.7 to 4.3 seconds Manual: 0.7 to 7.1 seconds
Luminance unit
cd/m
2
or fL
Luminance range
0.01 to 999,900 cd/m
2
0.1 to 9,999,000 cd/m
2
Accuracy*1
(Luminance) ±2% ± 1 digit
(Chromaticity)
±0.004 (5 cd/m
2
or more)
(Luminance) ±2% ± 1 digit
(Chromaticity)
±0.004 (50 cd/m
2
or more)
Repeatability*1
(Luminance) 0.2% + 1 digit
(Chromaticity)
0.001 (10 cd/m
2
or more)
(Chromaticity)
0.002 (5 cd/m
2
or more)
(Luminance) 0.2% + 1 digit
(Chromaticity)
0.001 (100 cd/m
2
or more)
(Chromaticity)
0.002 (50 cd/m
2
or more)
Calibration standard
Konica Minolta standard/user-specified standard switchable
User calibration
channels
10 channels
Data memory
1,000 data
External display
(Number of significant digits)
(Luminance) 4 digits (Max.)
(Chromaticity) 4 digits
Internal display
(Number of significant digits)
(Luminance) 4 digits (Max.)
Interface
USB2.0
Power
AA-size batteries (x2), USB bus power, or optional AC
adapter
Current consumption
When viewfinder display is lit: 70 mA average
Operation
temperature/
humidity range
0 to 40°C, relative humidity of 85% or less (at 35°C)
Storage temperature/
humidity range
0 to 45°C, relative humidity of 85% or less (at 35°C)
Size
71×214×154 mm
Weight
850 g (without batteries)
Standard accessories Lens Cap
Eyepiece ND Filter
Eyepiece Cap
AA-size batteries (x2)
Hard Case CS-A12
Wrist Strap CS-A13
USB Cable T-A15
Data Management Software CS-S20
Optional accessories
Close-Up Lens No. 153/135/122/110
CCD Camera Adapter CS-A14
Illuminance Adapter CS-A15
White Calibration Plate (for 45-0) CS-A20
AC Adapter AC-A305J/L/M
*1 Standard Illuminant A; Standard measurement distance; Measurement time setting: Auto
Standard accessories
Optional accessories
USB Cable (2 m)
T-A15
Lens Cap
Close-Up Lens
No.153, No.135,
No.122, No.110
Illuminance Adapter
CS-A15
White Calibration Plate
(for 45-0)
CS-A20
Hard Case
CS-A12
Data Management Software
CS-S20
PC
(
commercially available)
AC Adapter
AC-A305J/L/M
Chroma Meter
CS-150 / CS-160
Eyepiece
Cap
AA-size batteries (x2)
(commercially available)
Eyepiece
ND Filter
CCD Camera
Adapter
CS-A14
Wrist Strap
CS-A13
214
8.5
151
19
63
176
154.3
66
90.3
28.5
71
31
38
CS
Tripod screw hole
Measuring distance
reference plane
Size: 1/4 inch
depth: 6.5 mm