E-93
Connecting to an External Device
Outputting to a Printer
Connecting the instrument to a printer with a printer cable allows printing the measured data or the color
difference target data stored in the instrument’s memory. Two printing methods are available. One is to
print automatically each time measurement is taken (this is called “Auto Print”; measured data only), and
the other is to print the currently displayed measured data, pass/fail results and color difference target
data.
M
emo
If “SPECT.GRAPH” is selected as the display mode, the spectral reflectance data of displayed data in each screen;
the measured data in the measured data screen or the color difference target data in the <TARGET> screen, will be
printed. If “COLOR GRAPH” is selected, the color difference and absolute data will be printed. If “PASS/FAIL” is
selected, either “PASS” (or “FAIL”) or the color difference and absolute data will be printed.
N
ote
• When connecting the connectors, make sure that they are correctly oriented and fastened securely with screws.
• Before connecting, make sure that the power to both the instrument and printer is turned OFF.
• When disconnecting the cable, be sure to grip the plug or connector and pull it. Do not pull on or forcibly bend
the cable.
• Do not touch the connector terminals with your hands, allow them to get dirty or apply excessive force to them.
• Make sure that the cable has a sufficient amount of slack. Pulling the cable taut may cause connection failure
or wire breakage.
• The instrument and printer must be connected using a printer cable that is wired as shown in “Printer Cable
Wiring Diagram” (page E-92).
Communication Parameters
Item
Setting
Baud rate
9600 bps
Character bits
8 bits
Parity
None
Stop bits
1 bit
Flow control
RTS/CTS
Applicable Printers: DPU-H245AS-A03 *serial connection/metric screw type (Seiko Instruments
Inc.)
In addition to the above printer, printers that meet the following specifications can be used.
Item
Specifications
Number of print lines
27 or more lines, Data input: RS-232C
Data control
BUSY
Baud rate
9600
Character bits
8 bits
Parity
None
Stop bits
1 bit
Basic function
CR (0DH) carriage return
Summary of Contents for CM-2600d
Page 1: ...Spectrophotometer CM 2600d 2500d E Instruction Manual Es Manual de instrucciones...
Page 11: ...E 9 Chapter 1 Before Using the Instrument...
Page 27: ...E 25 Chapter 2 Preparation for Measurement...
Page 62: ...E 60 Setting Color Difference Tolerances...
Page 63: ...E 61 Chapter 3 Measurement...
Page 79: ...E 77 Chapter 4 Other Functions...
Page 108: ...E 106 TASK Mode...
Page 109: ...E 107 Chapter 5 Troubleshooting...
Page 114: ...E 112 Troubleshooting...
Page 115: ...E 113 Chapter 6 Appendix...
Page 123: ...E 121 Dimensions...
Page 126: ......
Page 127: ...E Es 2002 2005 KONICA MINOLTA SENSING INC AFMAGA 1 Printed in Japan 9222 1856 40...