Select the measurement area using the switch beforehand.
Three types of calibration can be performed with this instrument.
• Zero calibration : Only the amount of stray light is measured in advance in order to eliminate the effects of the stray light.
• White calibration : The reflectance is measured in advance using a known calibration plate in order to provide a reflectance scale.
Calibration Stage
Use a calibration stage with the same number as that printed on the nameplate of the instrument.
Newly purchased calibration stages will have a different number than that shown on the nameplate of the instrument. Make sure
the calibration plate ID displayed on the <Calibration> screen matches the calibration stage number.
Zero calibration
Since this instrument stores the data of the previous zero calibration, repeating zero calibration every time the instrument is
turned ON is not necessary. If, however, the measurement conditions change greatly or the instrument is not used for a long
period of time zero calibration must be performed before white calibration.
Memo
• The effects of stray light (i.e., light generated due to the flare characteristics of the optical system) will be
compensated for automatically by the zero calibration data.
• The amount of stray light may change because of dust or dirt that has collected in the optical system, humidity,
repeated operation, or vibration and shock exerted on the instrument. In this case, performing zero calibration
periodically is recommended.
Notes
• If the instrument is not used for long periods of time, the zero calibration data stored in the instrument may
be lost. If the data is lost, zero calibration must be performed again.
Operating Procedure
Start the procedure from the measurement screen.
1
Press [MENU], and then use [
◀
] or [
▶
] to display the
<Calibration> screen.
Notes
• The following screen prompting calibration is
displayed when the instrument starts up. If zero
calibration has not been performed, the cursor will
appear on “Calibration (Including Zero Cal.)”.
Otherwise, the cursor will appear on “Calibration
(Excluding Zero Cal.)”.
Zero calibration date
2019/11/19 08:44:31
White calibration date
2019/11/19 09:41:33
Calibration
Calibration
(Excluding Zero Cal.)
Calibration
(Including Zero Cal.)
Calibration
Summary of Contents for CM-23d
Page 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Page 30: ...27...
Page 82: ...79...
Page 111: ...108 Setting 3 MEMO...
Page 112: ...109...
Page 132: ...129...
Page 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Page 138: ...135...
Page 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Page 142: ...139 Dimensions Unit mm...
Page 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...