19
White Calibration
• If the white calibration plate becomes dirty, use a blower to blow off
any dust.
• If the white calibration plate does not get dirty easily, wipe it with a
soft cloth soaked in ethanol.
• When a part other than the calibration plate becomes dirty, lightly
wipe the dirt off with a cloth dampened with water or soapy water.
Notes
• Be careful not to scratch the calibration plate.
• Never use solvents such as thinner or naphtha.
• Scratches or dirt on the white calibration plate may
affect measurement values.
Target Mask
• Use a blower to blow off dirt or dust on the target mask.
• If the dirt on the outside surface of the target mask does not come off easily, wipe using a soft cloth soaked in ethanol.
Notes
Do not touch the painted surface of the integrating sphere.
Inside the Integrating Sphere
Use a blower to blow off dust or dirt inside the integrating sphere.
Notes
Do not touch the white-coated inner surface of the integrating sphere, wipe it with a cloth or put an object
inside it. If the target mask is dirty, and the dirt cannot be removed by a blower or the like, contact a KONICA
MINOLTA-authorized service facility.
Calibration plate
Summary of Contents for CM-23d
Page 1: ...Spectrophotometer CM 23d En Instruction Manual Please read before using the instrument...
Page 30: ...27...
Page 82: ...79...
Page 111: ...108 Setting 3 MEMO...
Page 112: ...109...
Page 132: ...129...
Page 133: ...130 Troubleshooting 5 Chapter 5 Chapter 5 Troubleshooting Message List 131 Troubleshooting 133...
Page 138: ...135...
Page 139: ...136 Appendix 6 Chapter 6 Chapter 6 Appendix Specifications 137 Dimensions 139...
Page 142: ...139 Dimensions Unit mm...
Page 151: ...En 9222 AC5J 31 2020 KONICA MINOLTA INC CACADK...