Operation and maintenance manual RollUP - A02 01-2021
Page
37
to
50
6
Maintenance and care
6.1 Cleaning and disinfection
•
After each use, the RollUP must be cleaned of foreign bodies by shaking it out, removing it
by hand or brushing it off.
•
Body fluids, etc. can be rinsed out under a water jet.
•
Steam jet must not be used.
•
Water temperature must not exceed 40°C.
•
Soiled textile components can be washed in clear fresh water, possibly with the addition of
a commercially available mild detergent.
•
Washing temperature not above 40°C.
•
Disinfection with commercially available alcohol-based agents. (Follow the application
instructions of the respective manufacturer!)
•
Disinfection with pH-neutral disinfectants based on quaternary ammonium compounds
with chlorhexidine. (Follow the application instructions of the respective manufacturer!)
•
After washing with detergent and / or disinfectant wash at least twice with clear fresh
water to remove all detergent residues.
•
Spray disinfection with alcohol-based disinfectants is permitted. (Follow the application
instructions of the respective manufacturer!)
6.1
Drying
•
Wet or damp components of the RollUP must be dried outdoors or in well-ventilated
rooms (normal room climate is a guide).
•
Do not expose to direct sunlight or direct heat radiation from heating sources.
•
Remove the RollUP complete with accessories from the transport backpack and dry it
completely unrolled.
•
After drying, the entire system must be subjected to a visual and functional check.