Volume Calibration
Calibration of the receiver volume is performed from the screen shown above.
1.
Put the empty receiver in place and select
Find Base
.
2.
After the level follower finds the base, select
Stand By
.
3.
Remove the receiver and use a class A pipette to deliver a known volume of
toluene into the receiver.
4.
Put the receiver in place and select
Track
.
5.
The level follower should move up to the level of the toluene in the receiver.
6.
Enter the volume of toluene that was pipetted into the receiver in the Calibration
Volume Field.
7.
When the level follower has stopped moving click on the
Get Counts
button.