K60094 Portable Heated Oil Test Centrifuge
Operation and Technical Manual
K60094-Manual
-
8-
6 Maintenance
WARNING.
Disconnect power to the unit before
servicing to avoid exposure to high voltages and/or
temperatures which may result in personal injury or
death. If you have any questions about maintaining
your equipment, then please do not hesitate to
contact the Koehler technical service department.
6.1 Routine Maintenance
Clean the unit after each test. Build-up of oil
residues can cause problems with the
operation and efficiency of the centrifuge.
Periodically, check the sample tube cushion
inside the carousel shields. Replace if worn
excessively.
ALWAYS
check
moving
parts,
both
mechanical and electrical, for wear and stress.
Replace if necessary.
6.2 Replacement Parts
Some instrument parts may need to be replaced.
When ordering replacement part(s), please provide
the model number, serial number, and product
shipment date of your equipment so that we can
ensure you will receive the proper replacement
part(s).
Part Number
Description
K60094-23014
Rotor Assembly
K60002-03043
Short Tube Holder
K60002-03058
Short Tube PTFE Cushion
230-012-001
Band Heater, 140W,
12VDC
275-600-011
Digital Temperature
Controller, 12-24V
265-000-007
Thermocouple, Type K
090-012-003
Relay, 40A, 12VDC SPST
278-002-001
Fuse, 2A
278-020-001
Fuse, 20A
288-012-003
Motor, 12VDC